Int J App Pharm, Vol 10, Issue 4, 2018, 43-50Original Article


RAPID ASSAY DEVELOPMENT OF DICLOFENAC SODIUM COATED TABLET ASSAY USING FTIR COMPARED TO HPLC METHOD

ILMA NUGRAHANI1, NOVENSIUS DILLEN1

1School of Pharmacy, Bandung Institute of Technology, Indonesia
Email: ilma_nugrahani@fa.itb.ac.id

Received: 06 Mar 2018, Revised and Accepted: 13 Apr 2018


ABSTRACT

Objective: A lot of coated tablet preparations of diclofenac have been marketed. This research aimed to develop and validate a quantitative analysis method for diclofenac sodium coated tablet using Fourier Transform Infrared (FTIR), which never reported.

Methods: The quantification was done by measuring the sample spectra, which then was converted into its derivative. Areas under the curve (AUC) of the derivative spectrums were plotted against the concentrations; corresponding to the calibration graphic. Then, the validation method was carried out by evaluating the accuracy, precision, linearity, range, limit of detection (LOD), and limit of quantification (LOQ).

Results: The results showed that diclofenac sodium had a specific peak within the wavenumber range of 1550-1605 cm-1. This area showed linearity to concentration within the range 0.1-1.0% w/w, with coefficient correlation of 0.9998. Recovery was found within 98-102% w/w. The intra and inter-day precision showed a coefficient of variance below 2%. The LOD and LOQ were 0.0127% and 0.0424% respectively. Further, a comparative study was performed, between this method and the compendia method using HPLC. The results showed that the measurement method using FTIR has an advantage in terms of time and cost.

Conclusion: Based on all data, it is concluded that FTIR can be used as a valid alternative method. It is faster and more cost-effective for diclofenac sodium coated tablet content determination, compared to the compendia method.

Keywords: Assay, Comparative study, Diclofenac sodium, FTIR, HPLC


INTRODUCTION

Methods for active compound analysis have developed rapidly, especially since the invention of various analytical instruments such as HPLC and gas chromatography. The development of quantitative analysis is driven by the need of more rapid, accurate and simple method of analysis. On the other side, the efficiency of cost should be considered; beside the realized of green pharmacy which is strong issued and recommended to be supported [1-3]. One of the developing methods prospectively is the determination level of an active substance by using FTIR instrument due to its free solvent usage. FTIR method has been developed and reported as a method of quantitative analysis for some active substances [4-7]. This method was explained to offer some advantages, such as simple sample preparation and minimum usage of organic solvents.

Diclofenac sodium is a nonsteroidal anti-inflammation drug (NSAID) with a mechanism of action inhibiting the cyclooxygenase enzyme, causing decreasing of prostaglandin synthesis [8, 9]. For some purposes, the diclofenac tablets are formulated in the coated form, such as for extended release or protect from degradation [10-15]. Film coating of tablets and other solid dosage also formulated to mask the unpleasant taste and odours. Furthermore, coated tablet is produced to improve the appearance of dosage forms; for ease of swallowing; to achieve colonic drug delivery and controlled drug delivery, among others. Many excipient are added to be the coating materials for pharmaceutical products [10-15].

Assay method for diclofenac sodium coated tablet preparation in the compendia is by using HPLC [8, 9, 16]. Some developments also have been reported to improve the performance [17-19]. One of the difficulties in the coated-dosage form analysis is the matrix's separation. In the common analysis, the matrices are extraction by appropriate solvents, if needed also supported by other steps of chromatography. In some practical cases, the coating excipient is not separated. It will increase the viscosity of the analyte solution. Then, that causes troubles of the instrument. Besides, prolongs the time consumption. Another method which has reported to quantify diclofenac is spectrophotometry UV/visible [20, 21], and voltammetry-GC [22]. However these methods also need solvent for extraction

In this study, FTIR instrument was proposed as a quantitative method for coated tablet of diclofenac sodium directly without extraction. Diclofenac sodium formula is C14H10Cl2NNaO2 with molecular weight of 318.129 g/mol. Diclofenac sodium structure as shown in fig. 1 has amine bond (N-H), carbonyl bond (C = O), and carbon double bond (C = C) which showing strong peaks in the IR spectrum [8, 9, 16, 23-25]. These bonds are predicted to have a correlation with the quantity of the measured substance so that it is potentially used to determine diclofenac sodium content in a sample.

Fig. 1: Diclofenac sodium chemical structure

The purpose of this study was to develop a more rapid and simple assay method for diclofenac sodium in the coated tablet using FTIR directly. It means, without extraction and separation physically. Furthermore, the parameters of validation resulted would be compared with the trial obtained by method from the compendia, completely with a short cost analysis.

MATERIALS AND METHODS

Material

The materials used in this study were: standard of diclofenac sodium (PT Pharos), KBr crystals (IR-spectra grade), acetone (Brataco, Indonesia), aqua-bidestilata, methanol pro HPLC, phosphate buffer pH 2.5, tablet matrix base, micropore 0.45 μm filter paper, and diclofenac sodium coated tablet preparations as samples.

Apparatus

The apparatus used in this study are: mortar, electronic scales, measuring glass, beaker glass, micro-pipette, FTIR (Jasco-4200 type A, Japan), KBr plate, pellet pressure tool (Jasco mini-press MP-1), mini-oven (Memmert), HPLC (Hitachi L-7000).

Methods

The study began with infrared spectrum analysis of diclofenac sodium in the range of 4000-400 cm-1 (mid-IR). From transmittance spectra, it was changed to absorption with the correlation: A =-log [T] [23-25]. Further, derivatization of the absorption spectrum was done; next the peak which clearest and can be used as the basis for making calibration curve was determined. The spectra selected must have the high value and specific to the analyte. Besides, it must not overlap with the other component’s spectra within the sample. The most important is, it should be linear over the concentration. Peak that has fulfilled the criteria was used to plot the calibration curve. The method was then validated to state that the method used has been appropriate to determine the levels of the active substance used. Validation criteria evaluated include the parameters of accuracy, precision, linearity, specificity, range, LOD and LOQ [16, 26-31]. The validated analysis method was next used to determine the level of diclofenac in the coated tablet from the market. The result was then compared with the amount declared on the label of the packaging’s preparation. This method is compared with methods that have been validated in the pharmacopoeia in terms of time, cost, and LOD/LOQ.

Development of diclofenac sodium analysis method

Spectrum measurement

Variations of standard concentrations of diclofenac sodium in KBR crystals were made in 10 different concentration within the range of 0.1-1% w/w. The mixture was weighed for 10 mg, then put into KBr plate and compressed using a press device. The results were measured at wavenumbers 4000-400 cm-1 and repeated three times.

Derivatization and standard wavenumber range determination

Derivatization was performed to increase the separation between the peaks, as same as the other spectroscopy methods when some spectrums are overlapped [32-36]. The obtained transmittance spectrum was changed into absorption then was derivatized to facilitate appropriate peak selection. The derivatization results from all concentrations were overlaid together. From the overlaid spectrum, peaks that have high molar absorption, specific, and linear to concentration was chosen. The peak which meets the criteria was then used as the base for plotting the calibration curve.

Validation of analysis method

Validation evaluation was conducted include parameter of accuracy, precision, linearity, specificity, range, LOQ, and LOD. The parameter's tests are based on guidance in the references [16, 26-31].

Specificity

The spectrum of the active substance, the matrices, the matrices added by the drug, and the sample were overlaid with each other. Range of wavenumbers where the peak present in the spectrum of the drug substance but not in the matrix spectrum was selected.

Linearity

A series concentration of standard diclofenac was prepared in KBr IR spectra-grade with concentrations of 0.1% to 1.0% of 10 units. Each concentration was measured using FTIR, then the spectrum was derivatized. The AUC value of the selected peak is measured. A curve between standard concentration and AUC value was plotted. The acceptance criterion for linearity is the correlation coefficient ≥ 0.999 and coefficient of variance for regression function ≤ 2%.

Range

The concentration range was used for the test of accuracy, precision, and linearity.

Accuracy

The test was done using standard addition method. One-sample tablet was grounded as blank. It is then added in a series of diclofenac sodium standard with the amount of 80%, 100%, and 120%, respectively. Diclofenac sodium 100% is equivalent to 50 mg of diclofenac sodium standard. The standards were then sampled and ground together with KBr IR spectra-grade with a ratio of 1:100. Each concentration was measured using FTIR and derivatized. The AUC value of the selected peak was measured, then the measured value content of each standard is determined using the calibration curve. Accuracy criteria are accepted if the recovered values of these three concentrations are within the range of 98-102%.

Precision

Determination used standard addition method. One sample tablet was ground as blank, after that added to the blanket of a 100% diclofenac sodium standard equivalent to 50 mg of diclofenac sodium. Standard was then sampled and ground together with KBr IR spectra-grade with a ratio of 1: 100. The mixture was measured using FTIR six times on the same day (intra-day precision) and repeated on three different days (inter-day precision). The AUC value of the selected peak was measured, then the measured value content of each standard was determined using the calibration curve. Precision criteria are accepted if the coefficient of variance value ≤ 2%.

LOD and LOQ

Value of LOD and LOQ could be determined using following equation [31]:

with

Determination of diclofenac sodium levels using FTIR

Ten of sample tablets were taken. Each was weighed, and the average was calculated. All the tablets were ground until homogeneous, next mixed in KBr crystal with concentration of 1% w/w. The levels of the active substances in the sample were measured using the calibration curve that had been made. The result of the determination of the content was then compared with the content declared on the packaging.

Determination of diclofenac sodium levels using HPLC

The method of diclofenac sodium content assay was adopted from the method written in the British Pharmacopoeia 2013 [9].

Chromatography system

Details of chromatographic system used in the experiment

Mobile phase: phosphate buffer pH 2.5: Methanol (20:80)

Stationary phase: C8 column Ø 3 mm length 15 cm

Flow rate: 0.5 ml/min

Detector: UV lamp 254 nm

Vol. of sample: 20 µl

Calibration curve plotting

Standard series of diclofenac sodium concentration was prepared in the mobile phase with concentrations of 10, 25, 50, 75, and 100 ppm, respectively. Each concentration was measured using HPLC, then the retention time and AUC values of each concentration were recorded. The curve between concentrations with AUC values was plotted to obtain standard calibration curve.

Measurement of coated tablet samples using HPLC

Ten of coated tablets as the sample were taken and crushed, then weighed amount of powder equivalent to 50 mg diclofenac sodium. The powder was dissolved in 70 ml pro-HPLC methanol, next filled with mobile phase up to 100 ml. The solution was then filtered and measured using HPLC. The AUC value of the diclofenac sodium peak was measured and used to determine the level of the active substance in the sample [9].

RESULTS AND DISCUSSION

Development of diclofenac sodium assay method

Firstly, the transmittance of infrared which pass the sample was recorded. Quantitative analysis of component in a sample can be measured using FTIR if there is a specific and suitable peak in the spectrum of the target component. The peaks selected for analysis should have high molar absorption. Furthermore, it does not overlap with other peaks of other components within the sample, symmetrical, and have linear correlation between absorbance and concentration. The absorption data was calculated automatically from transmittance by the instrument with the formula [23-25]:

A =-log T [. ]

A = Absorbance T = Transmittance

Measurement of levels using FTIR is based on Beer-Lambert's law, which says that the absorbance value of a light as it passes through the sample will be directly proportional to the thickness and concentration of the sample [23-25].

A = εcl

A = absorbance c = concentration

ε = molar absorptivity l = thickness

In Beer-Lambert's law, the plot curve between absorbance and concentration must be linear against 'εl' gradient and will pass through the origin point. Theoretically, to analyze an unknown component sample, some standard solutions with known concentrations need to be prepared, suitable peaks are selected, the absorption at the peak wave number is measured, and the calibration curve is plotted. The concentration of the sample component can then be determined using the sample absorbance value and the calibration curve that has been made [25].

So practically, the development of the method was begun with determining the peak that can be used as a standard for the assay. First, standard spectrum measurements were performed at concentrations of 0.1 to 1%. The obtained transmittance spectrum was then converted to absorbance, which next was derived to determine the corresponding peak as the base for plotting calibration curve [32, 33]. From the overlaid spectrum as shown in fig. 2 below, it is found that peaks in the range of 1550-1605 cm-1 tend to be linear. This peak corresponds to the stretching of the C = O bond [23-25].

Fig. 2: Overlay of derivative absorbance of standard spectrum of diclofenac sodium in concentration 0.1% to 1% w/w

FTIR assay method validation

Specificity

In the specificity test, spectrums of active substances, matrix, samples, and samples-added matrix were measured. This test aims to determine whether there is any disturbance caused by the matrix component of the sample. The spectrum resulted is not disturbed by the matrix (fig. 3).

Spectrums of the coated tablets and the uncoated tablets were measured. They were compared to determine whether the coating agent had effect or not on the diclofenac sodium spectrum.

From the overlay result as displayed in fig. 4, the spectrum of coated tablets and the spectrum of uncoated tablets did not have significant differences. So it explains that the coating agent did not affect the spectrum of the whole component of tablet.

Fig. 3: Overlay of spectrums of active substances, matrix, and samples-added matrix

Fig. 4: Overlay of transmittance spectrum between the coated tablet sample with un-coated tablet, sodium diclofenac bulk, and the mixture of sodium diclofenac with matrix

Linearity

Linearity is the ability of an analytical method to obtain measurements that are directly proportional to the concentration or number of analyte in a sample at a given range [26-31]. Linearity testing includes standard measurements of each concentration three times with re-sampling in every measurement. The measurement results are shown in table 1.

Table 1: Diclofenac sodium calibration curve data using derivative absorbance

Concentration (%) AUC (n=3) Average*
1 2 3
0.1 0.56577 0.51979 0.54471 0.54342±0.02302
0.2 1.07567 0.99456 1.01292 1.02772±0.04253
0.3 1.47897 1.48985 1.53707 1.50196±0.03089
0.4 2.03299 2.03112 2.01511 2.02641±0.00983
0.5 2.48560 2.57541 2.69375 2.58492±0.10440
0.6 3.07040 3.05048 3.04385 3.05491±0.01382
0.7 3.58413 3.57315 3.56369 3.57366±0.01023
0.8 4.07479 4.06845 4.09385 4.07903±0.01322
0.9 4.59470 4.58005 4.59168 4.58881±0.00774
1.0 5.13440 5.09514 5.14792 5.12582±0.02742

Note: AUC: area under the curve of absorbance, *All values are reported as mean±SD (n=3)

Fig. 5: Diclofenac sodium calibration curve

From test, a linear regression equation of diclofenac sodium was obtained (fig. 5). The regression equation was y = 5.101x+0.0051 with a correlation coefficient value equaled to 0.9998 and coefficient variance of regression function were 0.772%. It was concluded that the method has met the linearity criteria in the concentration range of 0.1% to 1% w/w.

Accuracy

Accuracy expresses the proximity between the value obtained from the measured result to the reference value or the real value [26-31]. Accuracy testing was done by standard addition method. The measurement results are shown in table 2.

Table 2: Diclofenac sodium accuracy test data

Added concentration AUC value Average of AUC value* AUC value theoretically %Recovery
80% 2.3685 2.3505±0.0064 2.3435 99.54%
2.3370
2.3684
2.3420
2.3355
2.3518
100% 2.6050 2.6028±0.0058 2.5985 99.69%
2.5811
2.5963
2.6126
2.6248
2.5975
120% 2.8653 2.8498±0.0058 2.8536 99.09%
2.8444
2.8454
2.8364
2.8746
2.8329

Note: AUC: area under the curve of absorbance spectrum, *All values are reported as mean±CV (n=6)

From the results of accuracy testing at three concentrations shown in table 2, the method has met the accuracy criteria because it has a recovery value between 98-102%.

Precision

Precision in the analytical method states the proximity of the obtained results (degree of scatter) from several sample measurements of a homogeneous sample under certain conditions [26-31]. Precision testing was performed using a standard method of addition with a measurement of six times. The precision level of the analytical method is determined by the value of the coefficient variance. There are two precision level tested: intra-day precision and inter-day precision.

Intra-day precision states the rate of repeatability of results under the same conditions of work within a narrow time span while inter-day precision states the rate of repeatability of results under various working conditions, such as different days or equipments.

Table 3: Diclofenac sodium precision test results

AUCs of day 1 AUCs of day 2 AUCs of day 3
2.60499 2.61607 2.59175
2.58106 2.60909 2.60347
2.59627 2.59752 2.60396
2.61255 2.60807 2.61566
2.62478 2.60544 2.61947
2.59752 2.59412 2.61095
Average 2.60286 2.60505 2.60754
CV 0.57657% 0.30863% 0.38296%

Note: AUC: area under the curve of absorbance spectrum; CV: coefficient variance

From the test results in table 3, it was concluded that the quantification method of diclofenac sodium with FTIR has met the criteria of intra-day and inter-day precision. The value of coefficient variance from six times consecutive measurement on the same day and on three different days were below 2%.

Range

The range is the interval between the lowest concentration and the highest concentration of the analyte in the sample which having sufficient valid parameters. The range values used generally estimated from 80% to 120% of the concentration value of the test compound. In this study, the concentration range of diclofenac sodium from 0.1% to 1.0% has met the parameters of accuracy, precision, and linearity [26-31].

LOD and LOQ

LOD is the lowest value that can be measured by a method of analysis without any certainty of the accuracy and precision of the value obtained. LOQ is the smallest quantity that can be accurately and precisely measured by an analytical method [26-31]. In the validation of this method, an equation was used to determine the value of LOD and LOQ. LOD was obtained at 0.0127% w/w, meanwhile the LOQ at 0.0424% w/w.

Tablet sample testing

The method assay by FTIR which has been validated, then used to measure the sample content of the coated tablets on the market. The sample was measured six times. Next the diclofenac sodium content was determined using the previously obtained regression equation from the calibration curve.

From the assay results in table 4, the sample of diclofenac sodium coated tablets met the criteria of the pharmacopoeia requirement. It states the range 95 to 105% of the levels declared in the packaging.

Diclofenac sodium testing using HPLC

The coated tablet commonly needs a lot of solvents and many steps of extraction to separate the analyte from its matrices more than uncoated preparation. The excipients can include glidants (flow aids), diluents, binders or granulating agents and lubricants to ensure efficient compressing; disintegrants to promote tablet break-up in the digestive tract. Moreover, the sweeteners or flavours are added to enhance the taste; meanwhile, pigments to make the tablets visually attractive [10-15]. Therefore, chromatography is usually used for assay to overcome the trouble of matrices influence [16, 17]. HPLC is the method to diclofenac sodium tablet assay in USP, British Pharmacopoeia and Indonesian Pharmacopoeia V [8, 9, 16]. In this study; the assay method using was adopted from British Pharmacopeia. This step was in order to compare compendia method with the developed method.

Table 4: Sample testing results using FTIR assay method

AUC value Measured content in KBr mixture (mg) Measured content in each tablet (mg)
1.3375 0.2612 50.1511
1.3405 0.2617 50.2644
1.3326 0.2602 49.9685
1.3305 0.2598 49.8902
1.3263 0.2590 49.7321
1.3289 0.2595 49.8274
Average* 49.9723±0.2013
Declared content 50 mg/tablet
% Recovery 99.94%

*All values are reported as mean±SD (n=6)

Table 5: Diclofenac sodium calibration curve data using HPLC

Concentration (ppm) AUC Average*
1 2 3
10 510571 509588 510084 510081.00±491.5069
25 1294832 1297245 1306301 1299459.33±6046.647
50 2671175 2671231 2676947 2673117.67±3316.418
75 4016318 4024867 4041067 4027417.33±12570.06
100 5301927 5333334 5335229 5323496.63±18703.82
125 6719323 6751552 6747041 6739305.33±17451.58

*All values are reported as mean±SD (n=3), The calibration curve was arranged based on data in table 5, which is displayed in fig. 6 as follows:

Fig. 6: Diclofenac sodium calibration curve using HPLC

Based on the curve in fig. 6, the LOD and LOQ was yielded the value LOD: 1.46 ppm and LOQ: 4.88 ppm.

Furthermore, the coated tablets of sodium diclofenac found from the market, were determined their content using this established method. This experiment yielded the results as follows (table 6).

Table 6: Diclofenac sodium coated tablet sample testing results using HPLC

AUC value Measured content (mg) % Recovery
2666366 50.06 100.11%
2685082 50.40 100.81%
2689407 50.48 100.97%
2693681 50.56 101.12%
2701469 50.71 101.41%
2701112 50.70 101.39%
Average

100.97%

0.48%

SD

The data shows that the branded sample coated tablet fulfilled the compendia requirement (BP: 95–102% w/w) [9].

Comparison of analysis methods

After the method of diclofenac sodium assay using FTIR was validated, a comparison of this newly developed method to the method contained in the compendia was done. The comparable factors include the time required during the analysis, the cost required, as well as the detection capability of each method. The differences were listed in table 7, 8, and 9. Table 7 explains time factor, meanwhile table 8 illustrates the cost, then table 10 is about the sensitivity which represented by LOD and LOQ.

Time factor

The result of time used calculation is listed in table VII.

Table 7: Comparison of FTIR with HPLC method in time factor

Compared factors FTIR HPLC
1-2 h 7-8 h
Preparation duration Simpler sample and standard preparation procedure, only including grinding and pressing Longer preparation time because of the more complex preparation procedure, such as the mobile phase preparation
Measuring duration ±10 s per sample ±10 min per sample

In terms of time listed in table 7, FTIR has advantages over HPLC with shorter preparation duration and sample measurement duration. The preparation procedure for the analytical method using FTIR is simpler because it only includes mixing with KBr and pressing. Preparation for HPLC tends to be longer and complicated because it requires the preparation of a mobile phase with a particular composition. In addition, the duration of measurement using the FTIR instrument is much faster than that of the HPLC. The ease of preparation and short duration of the work process became the prominent points of the FTIR method.

Cost factor

The cost factor comparison is displayed in table 8 as follows.

Table 8: Comparison of FTIR method with HPLC method in cost factor

FTIR
Material Amount needed Estimated cost
KBr IR-spectra grade 2 g Rp 50.000,00
Total Rp 50.000,00
HPLC
Material Amount needed Estimated cost
Aquadest 1 l Rp 35.000,00
Methanol pro HPLC 1 l Rp 150.000,00

Etc. (ex: filter paper,

buffer solution material, micropipet tip)

- Rp 20.000,00
Total Rp. 205.000,00

In terms of cost, the method of diclofenac sodium analysis using FTIR also has advantages with cheaper material prices. The analytical method with FTIR requires only KBr as a carrier material in the assay while HPLC method requires a relatively large volume of mobile phase for measuring the content level. In addition, the HPLC requires the column as a stationary phase which is expensive. The maintenance of HPLC instruments also requires a higher cost than FTIR instruments in the long term. FTIR is concluded as a more economical method of analysis.

LOD and LOQ

The comparison of LOD and LOQ between the developed method with the British compendia’s method (HPLC) is shown in table 10.

Table 10: Comparison of FTIR with HPLC method in detection capability

Factors FTIR HPLC
LOD

0.0127 %/

127 ppm

0.000146 %/

1.46 ppm

LOQ

0.0424%/

424 ppm

0.000448 %/

4.48 ppm


From detection capability’s data as listed in table 10 above, HPLC showed the more adequate LOD and LOQ. It means; the sensitivity of FTIR method is about 1/100 times of HPLC. However, FTIR can measure the higher concentration than HPLC. This eminence is due to the limited range of measurements of each instrument. FTIR can detect analyte in solid samples within the range of 0.1 to 1.0% with plate thickness of 0.1-0.2 mm. HPLC has a measurement capability in lower concentration range of up to 1 ppm, but unable to give exact measurements at higher concentration, example at 1000 ppm.

Furthermore, all validation parameters have shown the equal values compared with other methods referred [17-22]. The main differences are the range and LOD/lOQ; which in average FTIR is about 1/100 time lower to HPLC [17, 18]. Other reported method is voltammetry and gas chromatography methods with LOD/lOQ were determined as 4.8 and 0.15 μg/ml for LSV and GC–MS, respectively [22]. So, the developed LSV and GC-MS yield the almost similar limit values to HPLC methods, then FTIR is less sensitive compared. However, FTIR almost equal to the spectrophotometry UV reported [20, 21], which shown the LOQ in the range 5-25 μg/ml of the spectrum at 270-282 nm.

Considering the accuracy and precision, it proved that developed FTIR-AUC-derivative analysis is suitable to interchange with the other methods. The sensitivity also appropriate due to the dose of FTIR is match in the range to be quantified accurately. This instrument can produce the accurate and precision measurement for the distinctive dose of the active compound, by derivative and AUC calculation. The main advantage in this method is free of solvent. Besides, it shows the simplicity, less costly, and easy to conduct. Further, this method is proven more friendlier for the environment, support the green chemistry issue. Therefore, this method should be proposed as an economically alternative for content determination of the high-dose drug. As prediction, base on this result, is still appropriate for assay of an amount about 50 mg/tablet directly.

CONCLUSION

FTIR-AUC-derivative method is suitable to measure diclofenac sodium in the coated tablet matrix in the range of concentration 0.1 to 1.0% w/w with plate thickness of 0.1–0.2 mm. The LOD and LOQ are 0.0127% and 0.0424% w/w respectively. This method was proven suitable to determine the active compound in the coated tablet preparations in the dose 50 mg. As comparison, FTIR has the advantage in terms of time and cost than compendia’s HPLC method, but its sensitivity is lower.

ACKNOWLEDGMENT

This research was supported by Capacity Improvement Program, LPPM ITB, Bandung Institute of Technology, Indonesia. The authors would like to thank to PT Phapros Pharmaceutical Industry, Indonesia, for the diclofenac material support.

AUTHORS CONTRIBUTIONS

All the author have contributed equally

CONFLICT OF INTERESTS

Declared none

REFERENCES

  1. Namies ́nik J. Green analytical chemistry. J Separation Sci 200;24:151-3.

  2. Rozera R, Khatik GL, Gupta V. Virtual Labs: Greener and economic approach for learning the Pharma-Lab. Int J Green Pharm 2017;11(4, Suppl 2):663-70.

  3. International Pharmaceutical Federation. FIP STATEMENT OF POLICY Environmentally sustainable pharmacy practice: Green pharmacy. [Internet]. Buenos Aires, Argentina: FIP Bureau; 2016 [cited 2017 Sep 24]. Available from: http://www. fip. org/www/uploads/database_file. php?id=376andtable_id=.

  4. Pandey S, Pandey P, Tiwari G, Tiwari R, Rai AK. FTIR spectroscopy: a tool for quantitative analysis of ciprofloxacin in tablets. Indian J Pharm Sci 2012;74:86-90.

  5. Rakesh P, Charmi P, Rajesh P. Quantitative analytical applications of FTIR spectroscopy in pharmaceutical applied areas. J Adv Pharm Educ Res 2014;4:145-57.

  6. Sahoo S, Chakraborti CK, Mishra SC, Nanda UN, Naik S. FTIR and XRD investigations of some fluoroquinolones. Int J Pharm Pharm Sci 2011;3:165-­70.

  7. Bhongade B, Talath S, Dhaneswar SA. Validated method for the quantitation of ciprofloxacin hydrochloride using diffuse reflectance infrared fourier transform spectroscopy. Inter J Spec 2014:1-4.

  8. Ministry of Health of Republic of Indonesia. Pharmacopoeia Indonesia, 5th ed. Jakarta: Ministry of Health of Republic of Indonesia; 2014.

  9. The British Pharmacopoeia Commission, British Pharmacopoeia 2013., [Internet] United Kingdom: The British Pharmacopoeia Commission; 2013 [cited 2017 Sep 24]. Available from: http://www. uspbpep. com/bp2013/data/5420. asp.

  10. Hussan SD, S. Roychowdhury, P. Verma, V. Bhandari. A review on recent advances of enteric coating. IOSR J Pharm 2012;2:5-11.

  11. Ofori­kwakye K, Adom ENN, Kipo SL. Preparation and in vitro characteristics of tablet cores coated with albizia, albizia/khaya and albizia/HPMC films. Int J App Pharm 2009;1:22-9.

  12. Adedokun M, Nkata C. Optimized delivery of diclofenac sodium formulated in a sustained release Raphia Africana hydrocolloid matrix. Int J App Pharm 2018;10:109-14.

  13. Sundari PT, Gaud RS. Formulation and evaluation of alcohol resistant dosage forms of diclofenac sodium. Int J Pharm Pharm Sci 2013;5:854-7.

  14. Ahmeda SI, Mangamoorib LN, Raoc YM. Formulation and characterization of matrix and triple­layer matrix tablets for oral controlled drug delivery. Int J Pharm Pharm Sci 2010;2:137–43.

  15. Nawle RB, Tadvee AA. Formulation and evaluation of time-release compression coated tablet containing acebrophylline for chronotherapy of asthma. Int J Pharm Pharm Sci 2014;6:231-7.

  16. United States Pharmacopeia (USP) 34 (NF 29), Chapter<621>, Edition 2011.

  17. Atto AR. New method for determination of diclofenac sodium by High Performance Liquid Chromatography. Tikrit J Pharm Sci 2012;8:60-7.

  18. Ahmed SM, Raparla LP, Omer M. RP-HPLC method development and validation for simultaneous estimation of diclofenac sodium and serratiopeptidase in tablet dosage form. Int J Res Pharm Nano Sci 2015;4:10-8.

  19. Al-kamarany MA, Karbane ME, Belomaria M, Azougagh M, Hubert PH, Alanazi F, et al. Development and validation of LC method for the diclofenac sodium release determination using accuracy profile concept. Asian J Chem 2012;24:4393-400.

  20. Mali A, Jadhav S, Mane P, Tamboli A. Development and validation of UV spectrophotometric estimation of diclofenac sodium bulk and tablet dosage form using area under curve method. PharmaTutor Mag 2015;3:21-5.

  21. Naveed S. UV spectrophotometric assay of diclofenac sodium available brands. J Innovations Pharm Biol Sci 2014;1:92-6.

  22. Yilmaz B, Ciltas U. Determination of diclofenac in pharmaceutical preparations by voltammetry and gas chromatography methods. J Phar Anal 2015;5:153-60.[23] Rouessac F, Rouessac A, Chemical analysis: modern instrumentation methods and techniques. 2nd ed. West Sussex, England: John Wiley and Sons; 2007.

  23. Skoog DA, Crouch SR, Holler FJ. Principles of instrumental analysis. 6th ed. Belmont, Australia: Thomson Higher Education; 2007.

  24. Stuart B, Infrared spectroscopy: fundamentals and application. West Sussex, England: John Wiley and Sons; 2004.

  25. International Conference of Harmonization, Q2 Validation of analytical procedures: text and methodology international conference on harmonization of technical requirements for registration of pharmaceutical for human use, ICH Harmonized Tripartite Guideline, Canada, 2005.

  26. Ahuja S, Dong MW. Handbook of pharmaceutical analysis by HPLC. New York, United States: Elsevier Academic Press; 2005.

  27. Chan C. Analytical method validation and instrument performance verification. Toronto, Canada: John Wiley and Sons; 2004.

  28. Ehmer J, Miller JH. Method validation in pharmaceutical analysis: a guide to best practice. Weinheim, Germany: Wiley-VCH; 2005.

  29. Shabir GA. Step-by-Step analytical methods validation and protocol in the quality system compliance industry. In: Analytical Methods Validation. Institute of Validation Technology Duluth; 2006.

  30. Shrivastava A, Vipin BG. Methods for the determination of LOD and limit of quantitation of the analytical methods. Chron Young Sci 2011;29:21-5.

  31. Nugrahani I, Citra K, Determination of thiamine HCl (vitamin B1) and pyridoxine HCl (vitamin B6) content in tablet by FTIR. Int J Pharm Pharm Sci 2016;8:257-64.

  32. Nugrahani I, Musaddah MV. Development and validation analysis of acyclovir tablet content determination method using FTIR. Int J App Pharm 2016;8:43-7.

  33. Hoang VD, Ly DTH,Tho NH, Nguyen HMT. UV Spectrophotometric simultaneous determination of paracetamol and ibuprofen in combined tablets by derivative and wavelet transforms. Sci World J 2014;:1-13.

  34. Lotfy HM. Absorbance subtraction and amplitude modulation as novel spectrophotometric methods for the analysis of binary mixtures. Int J Pharm Pharm Sci 2013;6:735-41.

  35. Khveshchuk PF, Rudakova AV. Use of derivative spectrofluorometry in the analysis of ajmaline in drug dosage forms. Pharm Chem J 1995;29:796–98.