Int J App Pharm, Vol 10, Issue 6, 2018, 103-106Original Article


PENETRATION OF HYDROPHILIC SULFORHODAMINE B ACROSSTHE PORCINE CORNEAEX-VIVO

WANACHATCHAIYASANa*, SANGLY P.SRINIVASb, PATTRAVEENIAMPREMc, WAREE TIYABOONCHAIc,d,e

aDepartment of Optometry, Faculty of Allied Health Sciences, Naresuan University,Phitsanulok, Thailand,bSchool of Optometry, Indiana University, Bloomington, IN, United States,cDepartment of Pharmaceutical Technology, Faculty of Pharmaceutical Sciences,Naresuan University, Phitsanulok, Thailand,dThe Center of Excellence for Innovation inChemistry (PERCH-CIC), Commission on High-Education, Ministry of Education,Thailand,eThe Center of Excellence in Medical Biotechnology, Naresuan University, Phitsanulok,Thailand
Email: wanachatch@nu.ac.th

Received: 14 Jul 2018, Revised and Accepted: 04 Sep 2018


ABSTRACT

Objective: Sulforhodamine B (SRB) is a hydrophilic tracer whose fluorescence is unaffected by pH unlike that of carboxyfluorescein.Therefore, SRB may serve as a better tracer when there are significant changes in pH. Thus, in this study, the suitability of SRB to assess the barrier properties of the cellular layers of the corneawas examined using a custom-built confocal scanning micro-fluorometer(CSMF).

Methods: The dye solution (0.1% SRB) was prepared in PBS-Ca2+and three experiments were performed ex vivo using freshly isolated porcine eyes.First, we investigatedthe penetration of SRB across the endothelium by injection of the dye into the anterior chamber.Next, we measured the penetration of SRB across the epithelium after exposing the ocular surface to the dye. Finally,we examined the penetration after exposure to the dye with detergent (Tween 20) and exposure to the dye concomitant with microneedle injuries. The dye concentration profiles across the cornea were measured using CSMF.

Results: SRB penetratedthe corneal endotheliumreadilyinto the stroma following injection into the anterior chamber in a time-dependent manner. Despite accumulation in the stroma, SRB did not partition into the epithelium.In agreement with these findings, the dyedid not cross the epithelium after topical administration. Co-administration with Tween 20 and injury to the epithelium with microneedles, however, led to penetration of the dye into thestroma.

Conclusions: SRB is a hydrophilic dye that can be used as an alternative fluorescent tracer to assess the barrier functionof the cellular layers of the cornea.

Keywords: Sulforhodamine B, Cornea, Endothelium, Epithelium, Confocal scanning microfluorometry


INTRODUCTION

The cornea is the main refracting element of the eye. Therefore, its transparency is critical for acute vision. The transparency of the cornea is dependent on the stromal deturgescence (i.e., a state of dehydration of the stroma). The stromal deturgescenceis maintained by the barrier and fluid pump functions of the corneal endothelium[1]. The resistance to a persistent fluid leak that occurs from the anterior chamber into the stroma isthe barrier function of the endothelium. The mechanism that counterbalances the residual leak is the fluid pump activity. Thus, both the barrier and pump functions are obligatory for the stromal deturgescence. The pump function entails water movement by the transcellular route, and it is driven by local osmotic gradients induced by vectorial solute transport mechanisms of the endothelium. The barrier function, on the other hand, refers to passive fluid flux via the paracellular route. Thus, the barrier function isdetermined by the integrity of the tight junctions of the endothelium. Since the electrical methods are not suitablein vivo, theintegrity of the tight endothelial junctionsis frequentlymeasuredbypermeabilitytohydrophilicsolutes [2-5]. Previously, hydrophilic dyes such as sodium fluorescein (MW = 376; log P =-1.52)andits derivative carboxyfluorescein (MW = 378 g/mol, log P =-1.5)have been employed to characterize the barrier function of the corneal endothelium in a clinical setting using an ocular fluorometer [3,6,7].

Both fluorescein and carboxyfluorescein exhibit many useful characteristics as non-toxic polar solutes. They possess excitation and emission spectra in the visible range and exhibit high quantum yield. Fluorescein is the dye of choice for human studies as it is approved for use in clinical practice [8]. However, in animal studies, carboxyfluorescein is the dye of choicesince itis relativelymore hydrophilic than fluorescein. Because of higher hydrophilicity, the permeabilityof the endothelium to carboxyfluoresceinwould be a better measure of the integrity of the tight junctions [9]. However, thefluorescence spectra of carboxyfluoresceinare sensitive to pH[10]. Thus, measurements ofits fluorescence and hence the calculatedpermeabilityare confounded by changes in stromal pH.

Sulforhodamine B (SRB) is ahydrophilic dye (MW = 581; log P =-2.02; water solubility = 10 g/100 ml)[11] with absorption and emission peaks at 565 and 585 nm, respectively. More importantly, the fluorescence of SRB is unaffected by pH in the physiological range[12,13]. Therefore, SRB is a better solute for permeability measurements compared to fluorescein or carboxyfluorescein. Previously, the penetration of SRB intothe rabbit corneal epitheliumhas been reported[14]. Although SRB penetrated into thestroma,the dye wasnot found to stain the epithelial cells as observed by fluorescence microscopy [14]. In another study, SRB was found to penetrate intomouse corneas, which were exposed to toxic substances and had caused damage to the epithelium[15]. However, the penetration of SRB across the porcine cornea (which issimilar to the human cornea [16,17] has not been reported. Therefore, the objective of this study was to explore SRB as an alternative fluorescent dye to assess the barrier integrity of the epithelial and endothelial layers of the porcine cornea. We havemeasuredthe penetration of SRBwithporcine cornea ex vivo using a custom-made confocal scanning microfluorometer (CSMF) [18-21]. Our results indicate that SRB can be a superior dye for the assessment of permeabilityof the cellular barriers of the cornea.

MATERIALS AND METHODS

Materials

SRB and penicillin/streptomycin were obtained from Sigma Chemical Co (St. Louis, Mo, U.S.A). All other chemicals and solvents were of analytical grade.

Preparation of porcine eyeballs

Porcine eyeballs were collected from a local slaughterhouse and transported to the laboratory within 2 h immersed in ice-cold phosphate buffered saline (PBS with Ca2+;pH 7.4)containing 200 μg/mlof penicillin/streptomycin.Eyeballs,without anyinjury tothe epithelium,wereplaced in a moisture chamber at room temperatureprior to use in the permeability assay.

Confocal scanning microfluorometer(CSMF)

The CSMF, which has been described previously [20,21], was employed to assess the transcornealpenetration ofSRB. The instrument is equipped with a water-immersion objective (Zeiss 40x; 0.75 NA and wd = 1.2 mm) and measures transcorneal fluorescence profiles with a depth resolution of 7 µm. All depth scans were performed at 10 µm/s. The excitation was obtained byawhite light emitting diode(LED). The output of theLED (10W; XML series; Cree Inc) was modulated at 10 kHz andfiltered through an interference filter (565+10 nm). The output wascoupled to the circular end of a fiber optic bundle. At the other end of the bundle, the fibers were placed in a v-groove to form an excitation slit. The SRB fluorescence (>585 nm) and scattered light were collected through a parfocal slit positioned in the eyepiece. A photomultiplier tube (R928HA Hamamatsu) and a photodiode (S1226-5BQ; Hamamatsu), coupled to two independent lock-in amplifiers, detected the fluorescence and scattered light, respectively. As shown in fig. 1, all measurements were performed with eyeballs positioned underneath the objective on a motorized XYZ translation stage (Newport Inc., USA). The fluorescence/scatter scans were performed under computer control using custom-made Lab VIEW software (National Instruments).

Fig.1:Confocal scanning micro fluorometer employed for the measurements of transcorneal penetration of SRB [20,21]

Penetration of SRB

Three groups of experiments were performed to characterizethe transient penetration of SRB across the (a) endothelium, (b) cornea with and without epithelium, and (c) injured epithelium.For the latter experiments, the epithelium was injured with microneedles or by exposure to detergent. For all experiments, 0.1% SRB solutionwas prepared in PBS with Ca2+(pH 7.4). Each group of experiments was repeatedin at least six eyes.

For characterization of transient penetration of SRB across the endothelium, SRB solution (40 µl) was injected slowly using a 27-gauge needle into the anterior chamber. The injection was made ~ 2 mm posterior to the limbus.Theneedle tip did not touch the endothelium during the injection. The fluorescenceand scattersignals were recorded subsequently using the CSMF for up to~ 100 min.

For characterization of the transient penetration of SRB across the cornea with and without epithelium, 100 µl of SRB solution was placed at the center of the cornea for 0.5, 1, 4, and 12 h in a dark, moist chamber at room temperature. For some corneas, the epithelium was debrided using a sterile stainless-steel bladebefore exposure to the SRB solution. After each incubation period, the eyeballs were washed three times with PBS (containingCa2+; pH 7.4) and subsequently the fluorescence and scatter profiles were recorded using the CSMF as before.

For characterization of the transient penetration of SRB into the stroma of corneas with compromised epithelium,we employed Tween 20 to breakdown the tight junctions and independently,using microneedles. Thus, in some corneas, the epithelium was exposed to 100 µl of SRB solution containing 0.5% Tween 20 (dissolving in PBS with Ca2+pH 7.4) for 6-12 h.In other corneas, a 27-gauge needle was used to make severalmicroscopicinjuries to the epithelium before exposure to 100 µl of theSRB solution for 6 h. After incubation, the eyeballs were washed three times with PBS as above and then fluorescence and scatter profiles were recorded using the CSMF.

RESULTS

Penetration of SRB across the endothelium

The transcorneal profiles of SRB fluorescence from a typical experimentat various intervals between 30-100 min after injection of SRB into the anterior chamberare shown in fig. 2.The SRB fluorescence across the stroma increased in a time-dependent manner indicating significant permeability of the dye across the corneal endothelium and consequent accumulation in the stroma. Despitethe accumulation of the dye in the stroma to very high levels, it didnot subsequentlypenetrateinto the epithelium. As shown in fig. 2, a typical corneal scatter versus depth was taken at 10 min after injecting the dye. The peak heights between the scatter from the epithelium and endothelium can be taken as the thickness of the stroma, which was found to be about 650 µm (n = 6).

Fig.2: Penetration of SRB across the corneal endothelium after injection of the dye into the anterior chamber of with porcine eyesex vivo. The transcorneal fluorescence andscatter profiles were obtained 100 min after injection of SRB (0.1%). Results shown aresimilar to 6 other independent experiments

Penetration of SRB across the epithelium

The transcorneal profiles of SRB fluorescence with and without epithelium are shown in fig. 3 and fig. 4, respectively. As noted in fig. 3, there is no increase inSRB fluorescence within the corneal stromaeven after 12 h of exposure to SRB at the ocular surface. This indicates that SRB does not penetrate the epithelium. The apparent peak in SRB fluorescence at the epithelium in fig. 3 is due to adherent dye as the corneal surface could not be washed rigorously to prevent damage to the epithelium. In contrast to thefindings in fig. 3, in the case of the cornea without the epithelium, we found SRBtorapidly partition into the stroma within 30 min (fig. 4). Thus, the data in fig. 3 and fig. 4, taken together, indicate that the epithelium is a barrier to the penetration of the hydrophilic SRB.

The observations in fig. 3 and 4 have been further confirmed with models of injury to the corneal epithelium. As the first model,we damaged the tight junctions of the epithelium by exposure to0.5% Tween 20, a nonionic detergent. As shown in fig. 5, the transcorneal fluorescenceof SRB increased significantly in the presence of the detergent. In the second model, we caused mechanical injury to the epithelium using a microneedle. Similar to Tween 20, penetration of SRB also increased after the microneedle injury to the epithelium as shown in fig. 6.

Fig.3: Penetration of topical SRB with intact epithelium,the transcorneal fluorescence and scatter profiles 12 h after application of SRB (0.1%) to the epithelial surface.Results shown are similar to findings in 6other independent experiments

Fig.4: Penetration of topical SRB after removal of the corneal epithelium,the transcorneal fluorescence and scatter profiles 30 min after application of SRB on barestroma.Results shown are similar to findings in 6 other independent experiments

Fig.5:Penetration of SRB after disruption of the tight junctions of the epithelium,the transcorneal fluorescence and scatter profiles 12 h after application of SRB with Tween 20 (0.5%) at the epithelial surface,results shown are similar to findings in 6other independent experiments

Fig.6: Penetration of SRB after disruption of corneal epithelium using needles,the transcorneal fluorescence and scatter profiles 12 h.Results shown are similar to findings in 6other independent experiments

DISCUSSION

Mostophthalmic drugsare administered as drops on the ocular surface[22]. Hence, the studies on drug penetration across the cornea are vital to the rational design of ophthalmic drugs and their delivery strategies[23-25].The corneal epitheliumconsists of 5-7 layers of cells. The superficial two layersare thought tocontain a significant number of tight junctional strands, and hence form a barrier for hydrophilic solutes. On the other hand, the monolayer of corneal endothelium at the posterior surface of the cornea is very leaky. The stroma, which contains 80% water in a collagenous matrix,would be opposed to the accumulation of hydrophobic solutes but may form a depot for hydrophilic drugs. In this context, the permeability of SRB, a pH-insensitive hydrophilic molecule, into the porcine corneahas been evaluated usingthe custom-madeCSMF. The data not only highlight the cellular barriers for transcorneal penetration of drugsbut alsopredictive of the integrity of the cellular layers.

Our experiments have been performed with the freshly isolated porcine cornea, which formsanexcellent model,similar to rabbits,for studies on the transcorneal penetration of drugs and solutes[7,26]. The porcine cornea is structurally similar to the human cornea [17,27], although thicker than the human cornea[28]. The thickness of the porcine cornea isreported to be ~ 1010μm[29]. In our experiments, thethickness,measuredas the distance between peaks corresponding to the epithelium and endothelium from the scatter plots, was ~ 900 µm (ranged from 800 to 1000 µm; n = 12).

The penetrations of hydrophilic solutes into the corneal stroma are dependent on the integrity of the tight junctions of the corneal epithelium and endothelium. Thus, to assess the status of tight junctions, which is affected by inflammation or cell death, the permeability to hydrophilic dyes has been measured. For these investigations, an ideal dye would have pH-insensitive fluorescence properties along withnon-cytotoxicity to the corneal epithelium or endothelium. SRB is similar to carboxyfluorescein, which has long been used as a hydrophilic tracer forcharacterizingparacellular flux exclusively[30]. The data in fig. 3 and fig. 5 demonstrate that SRB penetrates the cornea after topical instillationexclusively through theparacellular pathway.As evident by data in fig.3 and 4 together,SRB does not penetrate the corneal epithelium. This is in contrast to the Rhodamine B, which is relatively lipophilic[31].In agreement with our findings in fig. 3 and 4, SRB partitioned very well into the stroma once the epithelium was removed (similar to sodium fluorescein that stains the cornea devoid of the epithelium[32]) (fig. 4) or when the integrity of the epithelium was disrupted by Tween 20 (fig. 5). The latter observation is reminiscent of the finding that Tween 20 (1%)induced an increase in the penetration of fluorescein into the human eyes[33]. The breakdown of the tightjunctions at the concentration of Tween 20 employed in this study has also been reported previously for Caco-2 cells[34]. The hydrophilic characteristics of the SRB are further demonstrated by the fact that the dyedoes not penetrate the epithelium, even after significant accumulation in the stroma(fig. 1).Finally, as expected, we found that the microneedle injury enhanced the permeability of SRB across the epithelium (fig. 6).

Collectively, our observations indicate that SRB can be used as an alternative hydrophilic tracer to assess the barrier functions of the corneal epithelium and endothelium.The characteristics of SRB penetration from the anterior chamber are similar to that of carboxyfluorescein as reported previously with the CSMF[1].

CONCLUSION

The findings of our study indicate that SRB could be employed as an alternative hydrophilic tracer to assess the barrier functions of the corneal epithelium and endothelium ex vivo.SRB is insensitive to pH, and hence, could be more advantageous as a tracer forassessment ofparacellular permeability.

ACKNOWLEDGMENT

The authors gratefully acknowledge financial support bytheNaresuan University Research Fund, Thailand (Grant number R2560C038).

AUTHORS CONTRIBUTIONS

All the authors have contributed equally

CONFLICTS OF INTERESTS

The authors declare that there is no conflict of interest

REFERENCES

  1. Srinivas SP. Dynamic regulation of barrier integrity of the corneal endothelium. Optom Vis Sci 2010;87:239-54.

  2. Srinivas SP. Cell signaling in regulation of the barrier integrity of the corneal endothelium. Exp Eye Res 2012;95:8-15.

  3. Srinivas SP, Goyal A, Talele DP, Mahadik S, Sudhir RR, Murthy PP, et al. Corneal epithelial permeability to fluorescein in humans by a multi-drop method. PLoS One 2018;13:e0198831.

  4. Nagataki S, Brubaker RF, Grotte DA. The diffusion of fluorescein in the stroma of rabbit cornea. Exp Eye Res 1983;36:765-71.

  5. Srinivasan B, Kolli AR, Esch MB, Abaci HE, Shuler ML, Hickman JJ. Teer measurement techniques for in vitro barrier model systems. J Lab Autom 2015;20:107-26.

  6. Araie M, Maurice D. The rate of diffusion of fluorophores through the corneal epithelium and stroma. Exp Eye Res 1987;44:73-87.

  7. Gupta C, Chauhan A, Srinivas SP. Penetration of fluorescein across the rabbit cornea from the endothelial surface. Pharm Res 2012;29:3325-34.

  8. Robertson TA, Bunel F, Roberts MS. Fluorescein derivatives in intravital fluorescence imaging. Cells 2013;2:591-606.

  9. Araie M. Carboxyfluorescein. A dye for evaluating the corneal endothelial barrier function in vivo. ExpEye Res 1986;42:141-50.

  10. Mordon S, Devoisselle JM, Maunoury V. In vivo ph measurement and imaging of tumor tissue using a ph-sensitive fluorescent probe (5,6-carboxyfluorescein): Instrumental and experimental studies. Photochem Photobiol 1994;60:274-9.

  11. Amato DV, Lee H, Werner JG, Weitz DA, Patton DL. Functional microcapsules via thiol-ene photopolymerization in droplet-based microfluidics. ACS Appl Mater Interfaces 2017;9:3288-93.

  12. Behrendorff N, Shukla A, Schwiening C, Thorn P. Local dynamic changes in confined extracellular environments within organs. Clin Exp Pharmacol Physiol 2009;36:1010-5.

  13. Schulz A, Hornig S, Liebert T, Birckner E, Heinze T, Mohr GJ. Evaluation of fluorescent polysaccharide nanoparticles for ph-sensing. Org Biomol Chem 2009;7:1884-9.

  14. Chodosh J, Dix RD, Howell RC, Stroop WG, Tseng SC. Staining characteristics and antiviral activity of sulforhodamine b and lissamine green b. Invest Ophth Vis Sci 1994;35:1046-58.

  15. Maurice D, Singh T. A permeability test for acute corneal toxicity. Toxicol Lett 1986;31:125-30.

  16. Hara H, Cooper DKC. Xenotransplantation–the future of corneal transplantation? Cornea 2011;30:371-8.

  17. Elsheikh A, Alhasso D, Rama P. Biomechanical properties of human and porcine corneas. Exp Eye Res 2008;86:783-90.

  18. Wang Y, Gabe K, Kompella UB, Barbieri B, Srinivas SP. Confocal scanning microfluorometer for transcorneal fluorescence lifetime measurements. Invest Ophth Vis Sci 2012;53:3093.

  19. Srinivas SP, Chaiyasan W, Niamprem P, Keefer K, Tiyaboonchai W, Kompella UB. Penetration of polar sulforhodamine b into the cornea. Invest Ophth Vis Sci 2015;56:4138.

  20. Maurice DM, Srinivas SP. Fluorometric measurement of light absorption by the rabbit cornea. Exp Eye Res 1994;58:409-13.

  21. Srinivas SP, Maurice DM. A microfluorometer for measuring diffusion of fluorophores across the cornea. IEEE Trans Biomed Eng 1992;39:1283-91.

  22. Gaudana R, Ananthula HK, Parenky A, Mitra AK. Ocular drug delivery. AAPS J 2010;12:348-60.

  23. Niamprem P, Srinivas SP, Tiyaboonchai W. Development and characterization of indomethacin-loaded mucoadhesive nanostructured lipid carriers for topical ocular delivery. Int J Appl Pharm2018;10:91-6.

  24. Chaiyasan W, Srinivas SP, Tiyaboonchai W. Development and characterization of topical ophthalmic formulations containing lutein-loaded mucoadhesive nanoparticles. Int J Pharm Pharm Sci 2016;8:261-6.

  25. Morsi N,Ghorab D,Refai H,Teba H.Preparation and evaluation of alginate/chitosan nanodispersions for ocular delivery. Int J Pharm Pharm Sci 2015;7:234-40.

  26. Chaiyasan W, Praputbut S, Kompella UB, Srinivas SP, Tiyaboonchai W. Penetration of mucoadhesive chitosan-dextran sulfate nanoparticles into the porcine cornea. Colloids Surf B 2017;149:288-96.

  27. Marquez-Curtis LA, McGann LE, Elliott JAW. Expansion and cryopreservation of porcine and human corneal endothelial cells. Cryobiology 2017;77:1-13.

  28. Sanchez I, Martin R, Ussa F, Fernandez Bueno I. The parameters of the porcine eyeball. Graefes Arch Clin Exp Ophthalmol 2011;249:475-82.

  29. Jay L, Brocas A, Singh K, Kieffer JC, Brunette I, Ozaki T. Determination of porcine corneal layers with a high spatial resolution by simultaneous second and third harmonic generation microscopy. Opt Express 2008;16:16284-93.

  30. Stiemke MM, McCartney MD, Cantu-Crouch D, Edelhauser HF. Maturation of the corneal endothelial tight junction. Invest Ophth Vis Sci 1991;32:2757-65.

  31. Gupta C, Chauhan A, Mutharasan R, Srinivas SP. Measurement and modeling of diffusion kinetics of a lipophilic molecule across rabbit cornea. Pharm Res 2010;27:699-711.

  32. Annear MJ, Petersen-Jones SM. Surgery of the ocular surface. In: Auer JA, Stick JA. editors. Chapter 57. Equine surgery (fourth edition). Saint Louis: W. B. Saunders; 2012. p. 770-92.

  33. Marsh RJ, Maurice DM. The influence of non-ionic detergents and other surfactants on human corneal permeability. Exp Eye Res 1971;11:43-8.

  34. Ujhelyi Z, Fenyvesi F, Varadi J, Feher P, Kiss T, Veszelka S, et al. Evaluation of cytotoxicity of surfactants used in self-micro emulsifying drug delivery systems and their effects on paracellular transport in caco-2 cell monolayer. Eur J Pharm Sci 2012;47:564-73.