Int J App Pharm, Vol 13, Issue 6, 2021, 210-215Original Article

QUANTIFICATION OF BILASTINE AND MONTELUKAST COMBINATION IN FORMULATIONS UTILIZING LIQUID CHROMATOGRAPHY: STABILITY STUDIES

KANCHARLA VIJAYALAKSHMI1*, BETHAPUDI SAMUEL ANAND ANDREWS2, BOLINENI NAGESWARA RAO3

1Quality Control Department, Divis Laboratories Limited, Hyderabad, India 508252, 2Department of Chemistry, Gitam Institute of Technology, GITAM University, Visakhapatnam, India 530045, 3Research and Development, Divis Laboratories Limited, Hyderabad, India 500018
Email: vijayalakshmikancharla.msc@gmail.com

Received: 27 Apr 2021, Revised and Accepted: 30 Aug 2021


ABSTRACT

Objective: We have developed a “stability-indicating RP-HPLC” procedure for the Bilastine (BLS) and montelukast (MTL) analysis of tablets.

Methods: The quantification of BLS and MTL combination was implemented utilising a Waters column (C18, 5 μm, 250 mm and 4.6 mm). Isocratic mobile phase had 60% volume KH2PO4 of 0.1M strength with pH 4.2 units and 40% volume methanol at a flow with 1.0 ml/min speed. UV detection at 232 nm was done to examine BLS and MTL. Stability experiments of BLS and MTL under distinctive environments of stress were also performed.

Results: The BLS and MTL were eluted at 1.810 min and 2.551 min, respectively. The responses were found to be linear for the concentration ranges of 10-30 µg/ml (BLS) and 5-15 µg/ml (MTL). Percent comparative standard deviance for precision was 0.331% (BLS) and 0.486% (MTL). Percent assay for accuracy was 98.96% (BLS) and 99.00% (MTL). The detection limit and quantitation limit measures for BLS were 0.018 µg/ml and 0.059 µg/ml, respectively, while for MTL it was 0.024 µg/ml and 0.081 µg/ml, respectively. Robustness studies authorized that the method is robust with percent comparative standard deviance of a highest 1.950%.

Conclusion: The developed “stability-indicating RP-HPLC” procedure for the BLS and MTL analysis is simple, sensitive, precise, specific and robust, making it appropriate to the assessment of BLS and MTL in a tablet formulation.

Keywords: Bilastine, Montelukast, Tablet formulation, Stability indicating, Analysis


INTRODUCTION

Nonsedating antihistamines are the first treatment option for the allergic rhinoconjunctivitis including urticaria, according to the existing recommendations [1, 2]. Bilastine (BLS) is not structurally relevant to many other antihistamines. BLS, like loratadine, desloratadine, even fexofenadine, falls in the piperidine grouping of antihistamines. BLS, as many other antihistamines, is also an inverse agonist for the H1 receptor. The in vitro tests have revealed that the BLS affinity with the H1 receptor is significant, but the affinity with 30 other checked receptors is very weak, or very little [3]. The in vivo tests have revealed histamine excited smooth muscle relaxation, endothelial permeability, bronchospasms, and microvascular extravasation were all decreased in the rats [4]. The suppression of histamine excited wheal and flare reaction behaviour in the skin, which was marked with BLS, was reported in vivo tests in the human populace.

From the findings of both comparative observations of montelukast (MLT) versus placebo and findings of MLT's preventive role on the bronchoconstriction occasioned by exercise or any other nonspecific triggers were reported, the first indications of MLT's efficacy in asthma were recorded [5]. MLT is a cysteinyl leukotriene receptor blocker that is intended to manage asthma as well as alleviate seasonal allergies signs. MLT works via attaching to a cysteinyl leukotriene receptor in the bronchial tubes and lungs and suppressing the operation of leukotriene D4 on it [6]. In mild-to-moderate asthmatics that are not taking inhaled corticosteroids, MLT improves symptoms, relief drug use, and pulmonary functioning as well as lowering the frequency of exacerbation and blood eosinophil quantities. Montelukast also outperformed long-acting beta2-agonists in preventing bronchoconstriction exacerbated by exercise [7].

The BLS and MTL structures were displayed in fig. 1. One publication resulted from a study of BLS and MTL absorbance grounded assays, suggesting BLS and MTL direct quantitative evaluation in the pharmaceutical dosage types [8]. BLS and MTL absorption at 214 nm and 281 nm, respectively, are used in their quantitation. For BLS and MTL analysis of tablets, no liquid chromatography-based approach has been put forward yet. In this investigation project, we developed a “stability-indicating RP-HPLC” method for the BLS and MTL analysis of tablets. We also studied the validated factors of “stability-indicating RP-HPLC” method proposed for the BLS and MTL analysis.

Fig. 1: BLS (bilastine) and MTL (montelukast) structures

MATERIALS AND METHODS

Chemicals

The BLS and MTL combination tablet kind used was Bilagio M (BLS 20 mg and MTL 10 mg, “Synokem Pharmaceuticals LTD, India”). “Rainbow Pharma Training Labs, India” provided the BLS and MTL reference samples. Methanol (Merck, India) and water (Milli Q water) utilized in “stability-indicating RP-HPLC” experiments were HPLC rating. NaOH, H2PO4, H2O2, KH2PO4 and HCl were all reagent rating from “Sd Fine Chemicals Ltd, India”.

Apparatus

The “stability-indicating RP-HPLC” studies were carried out on a Waters Corporation 2965 model high-performance liquid chromatography machine, which again was fitted with a PDA 2998 detector. The “stability-indicating RP-HPLC” quantification of the BLS and MTL combination was implemented, consuming a Waters column (C18, 5 μm, 250 mm and 4.6 mm).

BLS and MTL solutions

Stock BLS and MTL solution was freshly formulated in the diluent at quantities of 200 µg/ml (BLS) and 100 µg/ml (MTL). Stock BLS (200 µg/ml) and MTL (100 µg/ml) solution was next used to make fresh working BLS (20 µg/ml) and MTL (10 µg/ml) solution and calibration BLS (range: 10-30 µg/ml) and MTL (range: 5–15 µg/ml) solutions by diluting a proper volumes of stock BLS (200 µg/ml) and MTL (100 µg/ml) solution with diluent.

BLS and MTL analysing conditions

Mobile phase had 60% volume KH2PO4 of 0.1M strength with pH 4.2 units and 40% volume methanol at a flow with 1.0 ml/min speed. Temperature inside column, injector sample size and wavelength for the BLS and MTL enumeration was tuned at 25 °C, 10 μl and 232 nm, respectively. For the processing of BLS and MTL solutions, the mobile phase solvents blend was considered as a diluent.

BLS and MTL linearity curves

Prepared calibration BLS (range: 10-30 µg/ml) and MTL (range: 5–15 µg/ml) solutions by diluting proper volumes of stock BLS (200 µg/ml) and MTL (100 µg/ml) solution with the diluent. The BLS and MTL peak areas of the formulated solutions were reported at 232 nm under the proposed “stability-indicating RP-HPLC” method's conditions. The BLS and MTL peak areas recorded were next plotted against the related BLS and MTL concentrations. Thus, linearity curves for BLS and MTL were constructed which is followed by computation of regression equation.

Tablet analysis

Ten tablets of BLS and MTL commercial formulation, Bilagio M, was balanced and pounded. A portion of Bilagio M powder corresponding to BLS 20 mg and MTL 10 mg was correctly placed in a 100 ml flask and sonicated about 30 min with 50 ml diluent. Filtered this solution via membrane paper filter into a 100 ml another flask and finalized to 100 ml indication with the diluent. This is stock Bilagio M solution (200 µg/ml-BLS and 100 µg/ml-MTL). Stock Bilagio M solution was next used to make fresh working Bilagio M solution (20 µg/ml-BLS and 10 µg/ml-MTL). The BLS and MTL peak areas of the formulated solutions were reported at 232 nm under the proposed “stability indicating RP-HPLC” method's conditions. Using peak areas of BLS and MTL, their content in Bilagio M tablets were assessed using corresponding linearity curves or regression formulas.

BLS and MTL degradation investigation

Using Bilagio M solution, stress degradation of BLS and MTL under alkaline, acidic, photolytic, oxidative, and thermal environments were conducted [9].

Acidic condition

The stock Bilagio M solution (200 µg/ml-BLS and 100 µg/ml-MTL) was prepared. Ten millilitres each of stock Bilagio M solution and 0.1N HCl were correctly placed in a 100 ml flask and sonicated at ambient temperature for about 30 min. Filtered this solution via membrane paper filter into 100 ml another flask and finalized to 100 ml indication with diluent. The BLS and MTL peak areas of degraded specimen solution were reported at 232 nm under the proposed “stability-indicating RP-HPLC” method's conditions.

Alkaline condition

Prepared stock Bilagio M solution (200 µg/ml-BLS and 100 µg/ml-MTL). Sonicated ten millilitres each of stock Bilagio M solution and 0.1N NaOH that were correctly placed in a 100 ml flask at ambient temperature for around 30 min. Filtered this solution via membrane paper filter into 100 ml another flask and finalized to 100 ml indication with the diluent. Under the proposed “stability-indicating RP-HPLC” method's conditions, the BLS and MTL peak areas of degraded specimen solution were recorded at 232 nm.

Oxidative condition

The stock Bilagio M solution (200 µg/ml-BLS and 100 µg/ml-MTL) was prepared. Ten milliliters each of stock Bilagio M solution and 30% peroxide were correctly placed in a 100 ml flask and sonicated at ambient temperature for around 30 min. Filtered this solution via membrane paper filter into 100 ml another flask and finalized to 100 ml indication with the diluent. Under the proposed “stability-indicating RP-HPLC” method's conditions, the BLS and MTL peak areas of degraded specimen solution were recorded at 232 nm.

Thermal condition

Ten millilitres of stock Bilagio M solution (200 µg/ml-BLS and 100 µg/ml-MTL) was correctly placed in a 100 ml flask and exposed at 60 °C temperature for around 30 min. Filtered this solution via membrane paper filter into 100 ml another flask and finalized to 100 ml indication with the diluent. Under the proposed “stability-indicating RP-HPLC” method's conditions, the BLS and MTL peak areas of degraded specimen solution were recorded at 232 nm.

Photo condition

Exposed ten millilitres of stock Bilagio M solution (200 µg/ml-BLS and 100 µg/ml-MTL) that was correctly placed in a 100 ml flask to sunlight for nearby 24 hr. Filtered this solution via membrane paper filter into 100 ml another flask and finalized to 100 ml indication with the diluent. Under the proposed “stability-indicating RP-HPLC” method's conditions, the BLS and MTL peak areas of degraded specimen solution were recorded at 232 nm.

Validation

The validation factors, including selectivity, linearity, repeatability, accuracy, robustness and specificity, were checked that are agreed in ICH recommendation [10-12].

RESULTS

The major emphasis of this report is to establish a "stability implying RP-HPLC" system for determining BLS and MTL in tablets. Following multiple tentative trails, the following chromatographic settings were deemed to be desirable for determining BLS and MTL in tablets: Mobile phase-60% volume KH2PO4 of 0.1M strength with pH 4.2 units and 40% volume methanol at a flow with 1.0 ml/min speed, 25 °C of temperature inside column, 10 μl size of injection sample and 232 nm of wavelength for BLS and MTL enumeration. Chromatogram of BLS and MTL is made known in fig. 2.

Linearity

The calibration curves were obtained for BLS and MTL by injection (10 µl volume) of calibration BLS (range: 10-30 µg/ml) and MTL (range: 5–15 µg/ml) solutions. BLS and MTL area below their peaks were marked set against the corresponding BLS and MTL concentrations (fig. 3). The regression coefficient scores for the BLS and MTL and regression line formulas for BLS and MTL were:

For BLS-y = 123080 x–18048; R² score-0.9999

For MTL-y = 112576 x+8657.6; R² score-0.9997

Sensitivity

The “detection limit” (D-L) and “quantitation limit” (Q-L) are sensitivity parameters. These are computed utilising the ICH indorsed criteria [10]. The D-L and Q-L measures for BLS are 0.018 µg/ml and 0.059 µg/ml, respectively, while for MTL it was 0.024 µg/ml and 0.081 µg/ml, respectively.

Precision

Repeatability was evaluated with the working BLS (20 µg/ml) and MTL (10 µg/ml) solution injected (10 µl volume) six number of times. Mean, SD and % RSD for BLS and MTL peak areas acquired was evaluated (table 1).

Accuracy

Accuracy was evaluated with the working BLS (20 µg/ml) and MTL (10 µg/ml) solution injected (10 µl volume) six number of times. Mean, SD and % RSD for BLS and MTL content assay acquired was evaluated (table 1).

Fig. 2: Chromatogram of BLS (bilastine) and MTL (montelukast)

Fig. 3: BLS (bilastine) and MTL (montelukast) linearity curves

Table 1: BLS and MTL analysis precision and accuracy measures

Injection BLS Area MTL Area BLS % Assay MTL % Assay
I 2446909 1140264 99.16 99.05
II 2448991 1139441 99.24 98.98
III 2434640 1130618 98.66 98.21
IV 2436827 1147820 98.75 99.71
V 2451658 1138647 99.35 98.91
VI 2433083 1141501 98.60 99.16
Mean value (n) 2442018 1139715 98.96 99.00
SD value 8083.19 5534.498 0.328 0.481
RSD value 0.331 0.486 0.331 0.486

SD–standard deviation; R. SD–Relative standard deviation; n = 6 number of experiments

BLS and MTL degradation

BLS and MTL degradation under alkaline, acidic, photolytic, oxidative, and thermal environments revealed the results as follows: In an acidic environment, BLS and MTL were degraded by 10.08% and 9.54%, respectively. Alkaline environment degraded BLS and MTL at 7.59% and 6.36%, respectively. 4.73% of BLS and 8.26% of MTL were degraded in oxidative environments. When subjected to 60 °C temperature, 10.55% of BLS and 10.26% of MTL were degraded. In sunlight, BLS and MTL were degraded by 8.51% and 5.73%, respectively. The corresponding BLS and MTL degraded chromatograms are shown off in fig. 4. In acidic, photolytic, and thermal environments, four new peaks were detected in addition to the BLS and MTL peaks. While in alkaline and oxidative environments, three new peaks were detected in addition to the BLS and MTL peaks.

Fig. 4: BLS (bilastine) and MTL (montelukast) degradation investigation chromatograms

Table 2: BLS and MTL recovery measures

Added level µg/ml BLS added µg/ml BLS found % BLS Recovery Mean value (n) SD value RSD value
50% 9.900 9.88 99.81 99.36 0.408 0.411
9.900 9.83 99.27
9.900 9.80 99.01
100% 19.800 19.71 99.54 99.77 0.308 0.309
19.800 19.82 100.12
19.800 19.73 99.65
150% 29.700 29.66 99.88 100.00 0.161 0.161
29.700 29.75 100.18
29.700 29.68 99.93
Added level µg/ml MTL added µg/ml MTL found % MTL recovery Mean value (n) SD value RSD value
50% 4.950 4.96 100.23 100.32 0.145 0.144
4.950 4.97 100.49
4.950 4.96 100.25
100% 9.900 9.69 97.89 98.17 0.869 0.885
9.900 9.81 99.14
9.900 9.65 97.47
150% 14.850 14.84 99.95 99.70 0.255 0.256
14.850 14.81 99.70
14.850 14.77 99.44

SD–standard deviation; RSD–Relative standard deviation; n = 3 number of experiments

Recovery

A recognised amount of BLS (9.90, 19.80 and 29.70 μg/ml) and MTL (4.95, 9.90 and 14.85 μg/ml) comparable to make claims of 50%, 100% and 150% were included to the working Bilagio M solution (20 µg/ml-BLS and 10 µg/ml-MTL). The prepared specimen analysis was conceded three times under the proposed HPLC method's conditions. At every BLS and MTL concentration level, recoveries of BLS and MTL were gauged (table 2).

Robustness

The factors preferred for evaluating the robustness were: variation in wavelength (+2 nm, −2 nm), flow rate (+0.1 ml/min, − 0.1 ml/min), methanol proportion (+5% volume,-5% volume), pH (+0.1 unit,-0.1 unit) and column’s temperature (+, 2 °C, −2 °C). Robustness was inspected with the working BLS (20 µg/ml) and MTL (10 µg/ml) solution. The outcome of altered factors on the analysis of BLS and MTL was assessed in relations of Mean, SD and %RSD for BLS and MTL peak areas acquired (table 3).

System suitability

System suitability was inspected with the working BLS (20 µg/ml) and MTL (10 µg/ml) solution. The Mean, SD and %RSD for resolution, peak area, retention period, peak symmetry and theoretical plate number were determined (table 4) for BLS and MTL peaks conferring to ICH indorsed criteria [10].

Table 3: BLS and MTL robustness measures

Value BLS area Mean value (n) SD value RSD value MTL area Mean value (n) SD value RSD value
Methanol volume (%)
35 2504329 2454861 50075.69 1.040 1167392 1146396 20781.91 1.813
40 2456056 1145961
45 2404199 1125835
Flow speed (ml/min)
0.9 2409399 2456594 47467.29 1.932 1122730 1145361 22337.04 1.950
1.0 2504329 1167392
1.1 2456056 1145961
Wavelength detector (nm)
230 2508493 2462133 26219.65 1.056 1162504 1143421 20471.52 1.790
232 2481850 1121798
234 2456056 1145961
pH units
4.1 2456909 2457318 1509.78 0.061 1150264 1148555 2284.13 0.199
4.2 2458991 1149441
4.3 2456056 1145961
Temperature near column ( °C)
23 2456056 2459953 37853.77 1.539 1145961 1146587 21072.48 1.838
25 2424199 1125835
27 2499605 1167966

SD–standard deviation; R. SD–Relative standard deviation; n = 3 number of experiments

Table 4: BLS and MTL system suitability measures

Injection BLS Retention time BLS Area BLS peak plate count BLS peak tailing Resolution
I 1.808 2459329 7871 1.25 -
II 1.810 2458037 7980 1.24 -
III 1.809 2459859 7926 1.24 -
IV 1.810 2467652 7984 1.24 -
V 1.807 2456779 7921 1.23 -
Mean value (n) 1.809 2460331 7936.400 1.240 -
SD value 0.0013 4263.3018 46.8754 0.0071 -
RSD value 0.072 0.173 0.591 0.570
Injection MTL Retention time MTL Area MTL peak plate count MTL peak tailing Resolution
I 2.548 1144916 8500 1.20 4.71
II 2.550 1156005 8527 1.20 4.75
III 2.548 1145056 8530 1.20 4.73
IV 2.550 1146465 8596 1.19 4.78
V 2.548 1151907 8570 1.19 4.8
Mean value (n) 2.549 1148870 8544.600 1.196 4.754
S. D value 0.0011 4901.4368 38.0762 0.0055 0.0365
R. S. D value 0.043 0.427 0.446 0.458 0.767

SD–standard deviation; R. SD–Relative standard deviation; n = 6 number of experiments

DISCUSSION

High values of coefficient regression scores for the BLS and MTL determined indicated the worthy linearity of "stability implying RP-HPLC" system proposed for the BLS and MTL analysis [10, 13]. The very low measures of “detection limit” (D-L) and “quantitation limit” (Q-L) for the BLS and MTL determined indicated the desirable sensitivity of "stability implying RP-HPLC" system proposed for the BLS and MTL analysis [10, 14]. The enumerated recovery (table 2) achieves of BLS and MTL endorsing the selectivity besides non-interruption of the excipients in Bilagio M formulation [10, 14].

The 0.331% and 0.486% RSD measures for BLS and MTL, respectively proved the preciseness of the "stability implying RP-HPLC" system proposed. The 98.96% assay value measure for BLS and 99.00% assay value measures for the MTL proved accurateness of "stability implying RP-HPLC" system proposed [10, 15].

The method's stability suggesting the versatility has been evidenced by the sufficient segregation of all possible BLS and MTL degradation products (fig. 4) that were caused using alkaline, acidic, photolytic, oxidative, and thermal environments that are agreed in ICH recommendation [9,16-19]. BLS stability was in order of: Oxidative environment ˃ Alkaline environment ˃ Photo environment ˃ Acidic environment ˃ Thermal environment. MTL stability was in order of: Photo environment ˃ Alkaline environment ˃ Oxidative environment ˃ Acidic environment ˃ Thermal environment

Change in wavelength (+2 nm, −2 nm), flow rate (+0.1 ml/min, − 0.1 ml/min), methanol proportion (+5% volume,-5% volume), pH (+0.1 unit,-0.1 unit) and column’s temperature (+,2  °C, −2  °C) caused not beyond than 2% variance in the peak areas of BLS and MTL (table 3). This endorses robustness [10, 20]. The measures of the resolution, peak area, retention period, peak symmetry and theoretical plate number determined (table 4) for BLS and MTL proved the suitableness of the device for analysing the BLS and MTL combination [10, 21].

CONCLUSION

A "Stability implying RP-HPLC" method was suggested for the BLS and MTL determination in the Bilagio M formulation lacking excipients interference. The suggested approach has a faster run time. The method projected herein signifies the first effort for BLS and MTL determination in the dosage varieties. The system would also be applied to conduct standard quality management assessment of both BLS and MTL in the authorized pharmaceutical preparations that comprise both BLS and MTL.

ACKNOWLEDGEMENT

Authors wish to acknowledge GITAM Deemed to be University (Visakhapatnam, India), Rainbow Pharma Training Lab (Hyderabad, India), Dr. B. Nageswara rao, General Manager, Divis laboratories limited (Hyderabad, India) and Kolasani Srikanth, Assistant General Manager, Divis laboratories limited (Hyderabad, India) to carry out this work.

FUNDING

Nil

AUTHORS CONTRIBUTIONS

All authors have contributed equally.

CONFLICTS OF INTERESTS

Declared none

REFERENCES

1. Bousquet J, Khaltaev N, Cruz AA, Denburg J, Fokkens WJ, Togias A. Allergic Rhinitis and its Impact on Asthma (ARIA) 2008 update (in collaboration with the World Health Organization, GA(2)LEN and AllerGen). Allergy. 2008;63;Suppl 86:8-160. doi: 10.1111/j.1398-9995.2007.01620.x, PMID 18331513.

2. Bousquet J, Van Cauwenberge PV, Khaltaev N, Aria Workshop Group, World Health Organization. Allergic rhinitis and its impact on asthma. J Allergy Clin Immunol. 2001; 108(5);Suppl:S147-334. doi: 10.1067/mai.2001.118891, PMID 11707753.

3. Corcóstegui R, Labeaga L, Innerarity A, Berisa A, Orjales A. Preclinical pharmacology of bilastine, a new selective histamine H 1 receptor antagonist: receptor selectivity and in vitro antihistaminic activity. Drugs R D. 2005;6(6):371-84. doi: 10.2165/00126839-200506060-00005, PMID 16274260.

4. Corcóstegui R, Labeaga L, Innerárity A, Berisa A, Orjales A. In vivo pharmacological characterisation of bilastine, a potent and selective histamine H1 receptor antagonist. Drugs R D. 2006;7(4):219-31. doi: 10.2165/00126839-200607040-00002, PMID 16784247.

5. Paggiaro P, Bacci E. Montelukast in asthma: a review of its efficacy and place in therapy. Ther Adv Chronic Dis. 2011;2(1):47-58. doi: 10.1177/2040622310383343, PMID 23251741.

6. Baig S, Khan RA, Khan K, Rizvi N. Effectiveness and quality of life with montelukast in asthma- A double-blind, randomized control trial. Pak J Med Sci. 2019;35(3):731-36. doi: 10.12669/pjms.35.3.42, PMID 31258585.

7. Hon KL, Leung TF, Leung AK. Clinical effectiveness and safety of montelukast in asthma. What are the conclusions from clinical trials and meta-analyses? Drug Des Dev Ther. 2014;8:839-50. doi: 10.2147/DDDT.S39100, PMID 25061277.

8. Raj RM, Sankar ASK, Vetrichelvan T. Analytical method development and validation for simultaneous estimation of bilastine and montelukast sodium by UV spectrophotometry. World J Pharm Pharm Sci. 2020;10:680-7.

9. International Conference on Harmonization (ICH). Stability testing of new drug substances and products Q1A. Vol. R2. Geneva, Switzerland; 2003.

10. International Conference on Harmonization (ICH). Harmonized tripartite guideline validation of analytical procedures: text and methodology Q2. Vol. R1. Switzerland; 2005.

11. Ravichandran V, Shalini S, Sundaram KM, Rajak H. Validation of analytical methods-strategies and importance. Int J Pharm Pharm Sci. 2010;2:18-22.

12. Sharma S, Goyal S, Chauhan K. A review on analytical method development and validation. Int J App Pharm. 2018;10(6):8-15. doi: 10.22159/ijap.2018v10i6.28279.

13. Panchumarthy R, Navya ChN, Pravallika D, Sri DN. A review on step-by-step analytical method validation. IOSR J Pharm. 2015;5:7-19.

14. Locatelli M, Melucci D, Carlucci G, Locatelli C. Recent HPLC strategies to improve sensitivity and selectivity for the analysis of complex matrices. Instrum Sci Technol. 2012;40(2-3):112-37. doi: 10.1080/10739149.2011.651668.

15. Betz JM, Brown PN, Roman MC. Accuracy, precision, and reliability of chemical measurements in natural products research. Fitoterapia. 2011;82(1):44-52. doi: 10.1016/j.fitote.2010.09.011, PMID 20884340.

16. Soumia B, Fatima H, Saïd B, Bouchaib B, Souad T, Souad H, Ahmed B, Abdelmjid A. Statistical tools and approaches to validate analytical methods: methodology and practical examples. Int J Metrol Qual Eng. 2017;8:1-10.

17. Rode DM, Rao NN. A review on development and validation of stability-indicating HPLC methods for the analysis of acidic drugs. Int J Curr Pharm Sci. 2019;11:22-33. doi: 10.22159/ijcpr.2019v11i4.34939.

18. Rajasingam R, Sagineedu SR, Tan YH, Nalaiya J, Pichika MR. stress degradation studies and development of a validated RP-HPLC method for determination of tiagabine in the presence of its degradation products. Int J Pharm Pharm Sci. 2016;8:230-6.

19. Rao PV, Rao AL, Svum P. Development and validation of new stability-indicating reversed-phase high-performance liquid chromatography method for simultaneous determination of metformin hydrochloride and ertugliflozin in the bulk and pharmaceutical dosage form. Asian J Pharm Clin Res. 2019;12(1):235-40. doi: 10.22159/ajpcr.2019.v12i1.28938.

20. Ferreira SLC, Caires AO, Borges TdS, Lima AMDS, Silva LOB, dos Santos WNL. Robustness evaluation in analytical methods optimized using experimental designs. Microchem J. 2017;131:163-9. doi: 10.1016/j.microc.2016.12.004.

21. Epshtein NA. System suitability requirements for liquid chromatography methods: controlled parameters and their recommended values. [review]. Pharm Chem J. 2020;54(5): 518-25. doi: 10.1007/s11094-020-02231-w.