Int J Pharm Pharm Sci, Vol 8, Issue 2, 164-168Original Article


DEVELOPMENT AND VALIDATION OF A BIOANALYTICAL RP-HPLC METHOD FOR AZILSARTAN MEDOXOMIL WITH LIQUID-LIQUID EXTRACTION

SNEHAL KARPE, SANDEEP SONAWANE*, PRIYA RAHADE, SANJAY KSHIRSAGAR

MET’s Institute of Pharmacy, MET League of Colleges, Bhujbal Knowledge City, Adgaon, Nashik, Maharashtra, India
Email: sandeeps.iop@gmail.com

 Received: 13 Nov 2015 Revised and Accepted: 19 Dec 2015


ABSTRACT

Objective: There are many analytical methods available for estimation of Azilsartan medoxomil in biological samples and for pharmaceutical preparations. However, no specific RP-HPLC method with UV detection based on liquid-liquid extraction technique is available for estimation of Azilsartan medoxomil in human plasma.

Methods: A simple, rapid and accurate RP-HPLC with UV detection method was developed and validated as per US-FDA guidelines for the estimation of Azilsartan medoxomil in spiked human plasma using liquid-liquid extraction technique.

Results: Azilsartan medoxomil was well resolved from human plasma interference and internal standard (Aceclofenac) using C 18 (250 × 4.6 mm, 5 μ) column with methanol: 20 mm phosphate buffer (pH 3.0), (70: 30 %, v/v) as mobile phase, at a flow rate of 1 ml/min. The detection was performed at 249 nm. The calibration curve was found linear in the range of 500-16000 ng/ml. During calibration experiments, the heteroscedasticity was minimized by using weighted least square regression model with weighing factor 1/x². In accuracy and precision studies, intra-day and inter-day, % relative error was found between±15 and % RSD was less than 15 %. Stability experiments indicated that the drug remained stable after three freeze-thaw cycles.

Conclusion: A simple, rapid and accurate RP-HPLC method with UV detection was developed and validated for estimation of Azilsartan medoxomil based on liquid-liquid extraction technique. The developed method meets the requirements of US-FDA guidelines. Also the developed method does not require expensive chemicals and solvents and does not involve complex instrumentation, hence it is economic.

Keywords: Azilsartan medoxomil, Bioanalytical, Heteroscedasticity, Liquid-liquid extraction, RP-HPLC, Weighted regression.


INTRODUCTION

Azilsartan medoxomil chemically is (5–methyl–2–oxo-1,3–dioxol–4-yl) methyl 2–ethoxy–3-[[4-[2- (5–oxo-2H-1, 2, 4–oxadiazol–3-yl) phenyl] phenyl] methyl] benzimidazole–4-carboxylate [1]. It is Angiotensin II AT1 receptor blocker [2]. The structure of Azilsartan is given in [fig. 1].

Fig. 1: Structure of Azilsartan medoxomil

Literature survey revealed few methods for quantitative analysis of Azilsartan medoxomil in biological samples and in pharmaceutical formulations alone as well as in combination with other drugs. These include RP-HPLC method with PDA detection in human plasma by solid-phase extraction procedure [3], UPLC with MS/MS in beagle dog plasma with its application to pharmacokinetic studies by protein precipitation method [4], HPTLC method [5], determination of potential impurities in tablets by UPLC with UV detection [6], UV spectrophotometric method for determination of Azilsartan medoxomil in bulk and pharmaceutical dosage form [7]. Also, quantification methods of Azilsartan medoxomil in combination with Chlorthalidone include, LC with electrospray ionization operated in negative multiple reaction monitoring (MRM) mode in rat and human plasma by liquid-liquid extraction method [8], stability-indicating RP-HPLC method with UV detection [9] and spectrophotometric method based on first derivative spectra and spectro fluorometric methods[10].

Most of the earlier methods are based on the use of sophisticated instruments like LC-MS, UPLC-MS, which are not available in routine quality control laboratories and academic institute laboratories.

Therefore, the present work was undertaken with the objective of developing and validating a simple and rapid RP-HPLC with UV detection method for determination of Azilsartan medoxomil in human plasma based on liquid-liquid extraction technique which is cost effective and an efficient extraction technique. Also, the second objective of work was to suggest an approach in the selection of optimum calibration model.

MATERIALS AND METHODS

Materials and instrumentation

Pharmaceutical grade Azilsartan medoxomil and Aceclofenac used as an internal standard were provided as ex-gratis from Glenmark Pharmaceuticals, Sinnar, Nashik, India and Blue Cross Laboratories Ltd., Ambad, Nashik, India, respectively. Blank human plasma was provided as a gift sample from Dr. Vasantrao Pawar Medical College, Hospital and Research Centre, Nashik, India. Blank plasma was pooled by thoroughly mixing the plasma obtained from six different sources. Methanol used in the analysis was of HPLC grade, and all other chemicals were of analytical reagent grade, purchased from SD Fine Chemicals, Mumbai, India. The Durapore, 0.45 µ ´ 47 mm, membrane filter papers were purchased from Millipore (India) Pvt. Ltd., Bengaluru, India. Freshly prepared double distilled water used in the analysis was prepared from Borosil All Glass Double Distillation Assembly, purchased from Borosil, Mumbai, India.

Chromatographic analysis was carried out using an HPLC system consisting of pump PU-2080 Plus (JASCO Corporation, Tokyo, Japan) equipped with 100 µl Rheodyne loop injector (7725i) and detection was carried out on UV-2075 detector (JASCO Corporation, Tokyo, Japan) using Borwin Chromatography software (Version 1.50).

Preparation of standard stock solution and working standard solution for azilsartan medoxomil and aceclofenac

The stock solution of (1 mg/ml) of Azilsartan medoxomil was prepared in methanol and was further appropriately diluted with methanol to get six different working standard solutions with concentration 5, 10, 20, 40, 80 and 160µg/ml. Similarly, the stock solution (10 mg/ml) of Aceclofenac was prepared in methanol and appropriately diluted with methanol to get working standard solution of 100 µg/ml.

Preparation of calibration curve (CC) standards and quality control (QC) samples

Aliquots of 200 µl of pooled blank plasma were taken in stoppered glass tubes of capacity 20 ml. To this, 25 µl of 25 µg/ml methanolic standard stock solution of Azilsartan medoxomil (2500 ng) was added and to each tube, 25 µl of working standard solution of Aceclofenac (as internal standard) was added. The resulting solutions were vortex mixed for 1 min to get CC standards containing 500, 1000, 2000, 4000, 8000 and 16000 ng/ml of Azilsartan, respectively. The QC samples were similarly prepared to contain three concentrations [1500 ng/ml Low-Quality Control (LQC), 8000 ng/ml Middle-Quality Control (MQC) and 16000ng/ml High-Quality Control (HQC)].

Liquid-liquid extraction (LLE) experiments

Aliquots of pooled human plasma (200 µl) were taken in 20 ml stoppered glass test tubes. To each of these tubes 25 µl of 25 µg/ml methanolic solution of Azilsartan medoxomil and 25 µl of 100 µg/ml of working standard solution of Aceclofenac was added. To each tube, 3 ml of organic solvent was added, and the tubes were shaken in an inclined position on a reciprocating shaker at 100 strokes/min for 3 min. Further, these tubes were centrifuged at 3000 rpm for 10 min. The separated organic layer from each tube was transferred to separate glass tube and evaporated to dryness under a stream of nitrogen. The residue obtained upon evaporation to dryness was reconstituted with 250 µl of mobile phase and 100 µl was injected into HPLC system under optimal chromatographic conditions.

Chromatographic conditions

Chromatographic analysis was carried out on a C 18Phenomenex Hyperclone column (250 × 4.6 mm, 5 µm) with mobile phase consisting of methanol: 20 mm potassium phosphate buffer (pH 3.0), (70:30 %, v/v) at a flow rate of 1 ml/min. The detection was carried out at 249 nm.

Calibration runs

In the calibration experiments, 200 µl aliquots of all CC standards were analyzed in six replicates using optimized LLE method and appropriate chromatographic conditions. All calibration curves (CC) were analyzed in six replicates. Prior to analysis, each CC standard was mixed with 25 µl of 100 µg/ml methanolic solution of Aceclofenac (as internal standard). At the end of the calibration runs, the chromatograms of CC standards were processed to get the peak areas for Azilsartan medoxomil and Aceclofenac. For each CC standard, the area ratio of Azilsartan medoxomil and Aceclofenac was calculated.

Selection of calibration model and range

Data obtained from the run calibration experiments was subjected to unweighted and weighted least square regression analysis to generate the respective calibration equations [11]. In weighted regression, weighting factors (w) of 1/x and 1/x2 were used, where x is the concentration of the CC standards of Azilsartan medoxomil.

In order to select the best calibration model, each calibration model and equation was evaluated with respect to % Relative error (%RE), Residual plot and homogeneity of variance (homoscedasticity) in the linear range [12].

The area ratios for the CC standards were referred to the calibration equation to get the back-calculated concentrations (interpolated concentrations) for the each CC standards. The total % RE was calculated as the sum of % RE for all CC standards.

The predicted area ratios for CC standards were calculated by entering the nominal concentration of each CC standard into the calibration equation. A plot of residuals was constructed by plotting the differences between measured and predicted area ratios against nominal log concentration, and the scatter of residual was evaluated.

To evaluate homoscedasticity in the linear range, the variance of residuals at highest CC standard to the lowest CC standard was evaluated by means of one-way ANOVA.

The calibration model with minimum % RE, random scatter of points in the plot of residuals and no significant difference in one-way ANOVA was selected.

Validation studies

The developed method was validated as per US-FDA Guidance for Industry: Bioanalytical Method Validation (September 2013)[13]. Selectivity was studied at the lower limit of quantification (LLOQ) at 500ng/ml by comparing blank responses of plasma from six different sources with peak areas afforded by LLOQ samples. The Calibration curve standards were evaluated by preparing and analyzing CC standard solutions spiked with an internal standard for five days. The concentrations of each CC standard were back calculated using suggested calibration model and the deviation of the back-calculated concentrations from nominal values was studied and expressed as % nominal.

Precision and accuracy were studied by analyzing five bioanalytical batches over five days. Each batch consisted of one blank, all CC standards and five replicates of LQC, MQC and HQC samples. The calibration equation was determined for each batch from analysis of CC standards and was used to calculate the concentration of Azilsartan medoxomil in LQC, MQC and HQC samples. The within batch and between batch accuracy and precision was determined in terms of % RE and % RSD, respectively.

Stability of Azilsartan medoxomil in plasma was evaluated under various conditions viz. freeze-thaw cycles, stability at–20 °C for 30 d and stability at room temperature for 6 h. The amount of Azilsartan medoxomil in stability samples was found out and the % nominal and % RSD of the determinations were calculated.

RESULTS

When Azilsartan medoxomil and Aceclofenac were subjected to chromatographic analysis in mobile phases of different strengths and compositions, it was found that mobile phase consisting of methanol: 20 mm potassium phosphate buffer (pH 3.0), (70:30 %, v/v) gave adequate retention at a flow rate of 1 ml/min. The wavelength at which detection was carried out was 249 nm. The retention time for Azilsartan medoxomil was 5.53 min and for Aceclofenac, it was 9.61 min. various organic solvents like n-hexane, dichloromethane, diethyl ether and tert-Butyl methyl ether were tried in which good recovery was obtained with tert-Butyl methyl ether. Also, when aliquots of blank plasma were extracted with tert-Butyl methyl ether and chromatographed under mentioned chromatographic conditions, it was found that there were no significant interfering peaks at the retention times of Azilsartan medoxomil and Aceclofenac. Thus, it was concluded that tert-Butyl methyl ether could be further used as LLE solvent for Azilsartan medoxomil and Aceclofenac. The chromatogram of blank plasma extracted in tert-Butyl methyl ether is shown in [fig. 2(a)]and the chromatogram of Azilsartan medoxomil and Aceclofenac extracted in tert-Butyl methyl ether is shown in [fig. 2(b)]. The extraction recovery obtained for Azilsartan medoxomil and Aceclofenac was 58.25 % and 60.25 %, respectively.

During calibration experiments, when data obtained from [table 1]was subjected to unweighted and weighted linear regression with weighting factors 1/x and 1/x2;unweighted regression resulted in the equation Y = 0.0002x + 0.01099 and with 1/x and 1/x2weights, resulted in equation, Y = 0.0002x+0.0156 and Y = 0.0002x+0.0142, respectively. Each of the obtained linear regression equations was evaluated for % RE, random scatter and homoscedasticity for selection of appropriate calibration model [table 2]. From this, it was concluded that although all calibration equations gave random scatter residuals, the total % RE was minimal when weighted regression with weighting factor 1/x2 was applied. Further, when the Fcalculated values were compared with Ftabulated (a = 0.05), it became evident that a weighting factor of 1/x2 was suitable to homogenize the variance of the residuals. Thus, it was decided to adopt calibration model of weighted linear regression with weighing factor 1/x2 in the calibration range of 500 to 16000 ng/ml of drug.

Fig. 2(a): Chromatogram of blank plasma extracted in tert-Butyl methyl ether, 2(b): Chromatogram of Azilsartan medoxomil and Aceclofenac extracted in tert-Butyl methyl ether


Table 1: Area ratios from calibration experiments

CC

Amount of drug in ng/ml

Area ratio (mean±SD) (n=6)

1

500

0.1190±0.013

2

1000

0.2108±0.020

3

2000

0.4498±0.037

4

4000

0.8663±0.047

5

8000

1.5918±0.154

6

16000

3.3145±0.254


Table 2: Results of evaluation of various calibration models

Unweighted regression

Weighted regression (1/x)

Weighted regression (1/x2)

∑% RE

Nature of residuals plot

F5,5* value

∑% RE

Nature of residuals plot

F5,5** value

∑% RE

Nature of residuals plot

F5,5** value

36.18

Random scatter

349.207

-3.796-E

Random scatter

0.341

-8E-13

Random scatter

0.003

* F value calculated as ratio of variances at the extremes of the calibration range, ** F value calculated as ratio of variances of the weighted residuals


During validation studies, it was found that the peak areas for the lower limit of quantification (LLOQ) samples were more than five times the blank responses obtained using six different plasma sources which concluded that the method was deemed to be selective for an LLOQ of600 ng/ml. From [table 3], it was concluded that % nominal values of the back-calculated concentrations of CC standards were between 97-107 %, which were in acceptable limits.

Table 3: Standard curve parameters for Azilsartan medoxomil

CC

Nominal conc. (ng/ml)

 

Back calculated concentrations (ng/ml)

Mean

±SD

% R. S. D.

% Accuracy

1

2

3

4

5

6

1

500

499

431.5

599

449

537.5

575.5

515.2

67.53

13.10

103.05

2

1000

1037

927.5

876

848

1094

1062.5

974.2

103.6

10.64

97.42

3

2000

2136

1926

1981

2249

2236

2395.5

2154

176.7

8.20

107.7

4

4000

4326

3926

3991

4427

4499

4342.5

4251.9

236.5

5.56

106.2

5

8000

8035.5

6926

9207

7659

8020.5

7426.2

7879.0

770.6

9.78

98.48

6

16000

15926.5

15543.5

15927

16649

17170.6

18982

16699.8

1262.0

7.55

104.3

The evaluation of accuracy and precision showed that the intra-day % RE was between±15 %, while the % RSD was less than 15 %. The US-FDA Guidelines require that % RE be between±15 %, while % RSD should be less than 15 %. The results of precision and accuracy, as well as extraction recovery for Azilsartan medoxomil at LQC, MQC and HQC and for Aceclofenac, are presented in [table 4]. Further, the intermediate precision of the method was determined by using a one-way ANOVA. For each QC level, within mean square and between mean square values were determined. The total variance was taken as a sum of within and between mean squares and the standard deviation was determined as a square root of total variance and F value was determined. The Fcalculated was found less than Ftabulated (a = 0.05), [table 5] indicates that there is no significant difference between intra-day and inter-day precision.

Table 4: Results of accuracy and precision studies

Intraday n= 5

Interday n = 5

QC Level

Conc in ng/ml

Mean conc found ng/ml

%RE

%RSD

Mean conc found ng/ml

%RE

%RSD

% Recovery

LQC

1500

1513.2

0.82

4.64

1563.32

5.27

5.15

41.08

MQC

8000

8003.6

0.03

3.12

8378.5

4.70

4.99

58.4846

HQC

16000

17366.5

8.54

2.9

16970.1

6.06

4.54

46.88

IS

-

-

-

-

-

-

-

60.25


Table 5: Results of one-way ANOVA at each QC level

QC Level

Source

Sum of squares

Df

Mean squares

Total Variance

±SD

F value

LQC

Within run

168784

20

8439.2

12435.4

111.5

3.8

Between run

129666

4

32417

MQC

Within run

4586655

20

229333

3911007

625.3

5.2

Between run

4799763

4

1199941

HQC

Within run

1.490Eto7

20

745214

1230856

1109.4

4.9

Between run

1.463Eto7

4

3659064


Table 6: Results of stability studies for Azilsartan medoxomil

QC Level

Stability at RT

Stability at-20°C

Freeze-thaw stability

 

%Nominal

% RSD

%Nominal

% RSD

%Nominal

% RSD

LQC

102.21

3.37

102.3

3.85

101.26

4.9

HQC

101.73

3.22

100.6

5.66

100.23

6.92

The results of stability evaluation of Azilsartan medoxomil are presented in [table 6]. Analysis cycles viz. three freeze-thaw cycles, stability at-20 °C for 30 d and stability at room temperature for 6 h indicated that Azilsartan medoxomil was stable in human plasma under these conditions. The developed method is sensitive and convenient for use. The developed method does not require expensive chemicals and solvents and does not involve complex instrumentation or sample preparation methods and hence it is simple and economical as compared to previously reported methods.

CONCLUSION

In this report, a simple, rapid, selective and accurate HPLC-UV method was described for the quantification of Azilsartan medoxomil in spiked human plasma using liquid-liquid extraction. The developed bioanalytical method is capable of quantifying Azilsartan medoxomil from spiked human plasma in the concentration range of 500–16000 ng/ml. The method meets the requirements of the US-FDA guidelines.

ACKNOWLEDGEMENT

Authors are thankful to the management and trustees of Mumbai Educational Trust’s Bhujbal Knowledge City, Nashik, for providing necessary chemicals and analytical facilities; Glenmark Pharmaceutical Ltd. Sinnar, Nashik and Blue Cross Laboratories Ltd., Ambad, Nashik for providing Azilsartan medoxomil and Aceclofenac as ex-gratis, respectively. Authors are also thankful to Dr. Vasantrao Pawar Medical College, Hospital and Research Centre, Nashik for providing human plasma samples.

CONFLICT OF INTERESTS

Declared none

REFERENCES

  1. National Center for Biotechnology Information. PubChem Compound Database; CID=11238823. Available from: https://pubchem.ncbi.nlm.nih.gov/compound/11238823. [Last accessed on 10 Oct 2015].
  2. Ojima M, Igata H, Tanaka M, Sakamoto H, Kuroita T, Kohara Y, et al. In antagonistic properties of a new angiotensin type 1 receptor blocker, azilsartan in receptor binding and functional studies. J Pharmacol Exp Ther 2011;336:801-8.
  3. Vekariya PP, Joshi HS. Development and validation of RP-HPLC method for azilsartan medoxomil potassium quantitation in human plasma by solid phase extraction procedure. ISRN Spectroscopy; 2013. doi: 10.1155/2013/572170. [Article in Press]
  4. Cheng G, Junfeng W, Yinghua S, Dawei D, Zhong L, Meng Z, et al. UPLC-MS/MS for the determination of Azilsartan in beagle dog plasma and its application in a pharmacokinetics study. Asian J Pharm Sci 2014;10:247-53.
  5. Gandhi SV, Mittal PS, Pahade AR, Rege SW. Development and validation of stability indicating HPTLC method for estimation of Azilsartan medoxomil. Int J Pharm Sci 2014;6,224-32.
  6. Mantena BPV, Sumathi VR, Appa Rao KMCh, Ramakrishna K, Reddy RS. Method development and validation for the determination of potential impurities present in azilsartan medoxomil tablets by reverse phase-ultra performance liquid chromatography. Anal Chem Lett 2014;4:287-301.
  7. Raja G, Nagaraju CH, Sreenivasulu B, Sreenivas N, Korupolu RB. New simple UV spectrophotometric method for determination of azilsartan medoxomil in bulk and pharmaceutical dosage forms. Int J Res Pharm Biomed Sci 2013;4:1133-7.
  8. Rachumallu R, Puttrevua SK, Bhateriaa M, Balab V, Sharmab VL, Bhattaa RS. Simultaneous determination of Azilsartan and Chlorthalidone in rat and human plasma by liquid chromatography-electrospray tandem mass spectrometry. J Chromatogr B: Anal Technol Biomed Life Sci 2015;990:185-97.
  9. Naazneen S, Sridevi A. Stability indicating RP-HPLC method for the simultaneous estimation of azilsartan medoxomil and chlorthalidone in solid dosage forms. Int J Pharm Pharm Sci 2014;6:236-43.
  10. Walid ME, Elkady FB, El-Zaher A, El-Bagary RI, Patonay G. Spectrophotometric and spectrofluorometric studies on azilsartan medoxomil and chlorthalidone to be utilized in their determination in pharmaceuticals. Anal Chem Insights 2014;9:33–40.
  11. Nagaraja NV, Paliwal JK, Gupta RC. Choosing the calibration model in assay validation. J Pharm Biomed Anal 2012;20:433-8.
  12. Bolton S, Bon C. Linear regression and correlation In: Pharmaceutical Statistics Practical and Clinical Applications. Vol. 203. Informa Healthcare USA, Inc; 2010. p. 147-81.
  13. US Food and Drug Administration, Center for Drug Evaluation and Research, Center for Veterinary Medicine, Guidance for industry-Bioanalytical method validation; 2013. Available from: http://www.fda.gov/downloads/drugs/ guidance-compliance regulatory information/guidances/ucm368107.