COMPARISON OF SURFACE RADIATION DOSE TO THE GONADS BY RADIOGRAPHIC EXAMINATION OF THE LUMBAR SPINE USING COMPUTED RADIOGRAPHY AND DIRECT DIGITAL RADIOGRAPHY

Authors

  • Priyanka Department of Medical Imaging Technology, Manipal School of Allied Health Sciences, Manipal University, Manipal, Karnataka, India. http://orcid.org/0000-0003-2682-158X
  • Rahul P Kotian Department of Medical Imaging Technology, Manipal School of Allied Health Sciences, Manipal University, Manipal, Karnataka, India.
  • Nitika C Panakkal Department of Medical Imaging Technology, Manipal School of Allied Health Sciences, Manipal University, Manipal, Karnataka, India.

DOI:

https://doi.org/10.22159/ajpcr.2017.v10i12.21029

Keywords:

Surface radiation dose, Computed radiography, Direct digital radiography, Lumbar spine

Abstract

Objectives: The objectives are aimed to study the comparison of SRD to the gonads by radiographic examination of the lumbar spine using computed radiography (CR) and direct digital radiography (DR) and to evaluate dose differences according to gender.

Methods: A total of 120 subjects with an equal number of males and females referred for lumbar spine anteroposterior (AP) and lateral was included in the study. Sixty patients had undergone X-ray in CR X-ray unit and 60 in DR X-ray unit. SRD to gonads from a radiographic examination of the lumbar spine was measured in CR and DR using Mult-O-Meter, and obtained value of SRD was in μGy. Statistical analysis was performed using Statistical Package for the Social Sciences. Data were represented as a median and interquartile range. Mann–Whitney U-test was used for the comparison of SRD to gonads. Two-way analysis of variance (ANOVA) test was used to find out the statistically significant difference in SRD to the gonads according to gender from radiography of lumbar spine taken using CR and direct DR.

Results: There was a statistically significant difference in SRD to gonads from radiography of lumbar spine AP and lateral taken using CR and direct DR (p<0.001). There was no statistically significant difference in SRD to the gonads in males and females from radiography of lumbar spine AP (p=0.577) and lateral (p=0.164) taken using CR and direct DR.

Conclusion: It was found that SRD to gonads from lumbar spine AP was 54% lower in direct DR and SRD to gonads from lumbar spine lateral was 68% lower in direct DR than CR.

Downloads

Download data is not yet available.

Author Biography

Priyanka, Department of Medical Imaging Technology, Manipal School of Allied Health Sciences, Manipal University, Manipal, Karnataka, India.

SENIOR SCALE ASSISTANT PROFESSOR

References

Maher C, Underwood M, Buchbinder R. Non-specific low back pain. Lancet 2017;389(10070):736-47.

Tripathy A, Adiga S, Shah H, Shanbhag TV, Kumar MD. A retrospective study of clinical profile and drug prescribing pattern in osteoporosis in a tertiary care hospital. Int J Pharm Pharm Sci 2015;7(10):390-3.

Oborska-Kumaszynska D, Wisniewska-Kubka S. Analog and digital systems of imaging in roentgenodiagnostics. Pol J Radiol 2010;75(2):73-81.

Gupta M, Dahiya J, Marwaha RK, Dureja H. Therapies in cancer treatment: An Overview. Int J Pharm Pharm Sci 2015;7(4):1-9.

Ogilvy-Stuart AL, Shalet SM. Effect of radiation on the human reproductive system. Environ Health Perspect 1993;101 Suppl 2:109-6.

Parry RA, Glaze SA, Archer BR. The AAPM/RSNA physics tutorial for residents. Typical patient radiation doses in diagnostic radiology. Radiographics 1999;19(5):1289-302.

Suliman II, Habbani FI. Effective dose Calculations in Conventional Diagnostic X-Ray Examinations for Patients in Major Sudanese Hospitals. World Congress on Medical Physics and Biomedical Engineering; 2006. p. 208791.

Compagnone G, Baleni CC, Pagan L, Calzolaio FL, Barozzi L, Bergamini C. Comparison of radiation doses to patients undergoing standard radiographic examinations with conventional screen-film radiography, computed radiography and direct digital radiography. Br J Radiol 2006;79(947):899-904.

Panicker TM, Angelina JT, Korath MK, Mohandas K, Jagadeesan K, Jagadeesan K. Entrance skin dose estimation in X-ray Lumbar Spine lateral procedure : Conventional Vs Digital X-ray units . Pilot Study 2013;26(4):219-20.

Toosi MB, Nazery M, Zare H. Application of dose-area product meter to measure organ and effective dose in diagnostic radiology at two hospitals, Mashhad, Iran. In: 11th International Congress of the International Radiation Protection Association, Madrid; 2004. p. 1-10.

Gyekye PK, Simon A, Geoffrey ER, Johnson Y, Stephen I, Engmann CK, et al. Radiation dose estimation of patients undergoing lumbar spine radiography. J Med Phys 2013;38(4):185-8.

Kotian RP, Priyanka, Panakkal NC. A pilot study to compare surface radiation dose to the gonads by radiographic examination of the lumbar spine using computed radiography and direct digital radiography. Int J Pharm Biol Sci 2016;7(3):499-502.

Published

01-12-2017

How to Cite

Priyanka, R. P. Kotian, and N. C. Panakkal. “COMPARISON OF SURFACE RADIATION DOSE TO THE GONADS BY RADIOGRAPHIC EXAMINATION OF THE LUMBAR SPINE USING COMPUTED RADIOGRAPHY AND DIRECT DIGITAL RADIOGRAPHY”. Asian Journal of Pharmaceutical and Clinical Research, vol. 10, no. 12, Dec. 2017, pp. 52-55, doi:10.22159/ajpcr.2017.v10i12.21029.

Issue

Section

Original Article(s)