A VALIDATED RP-HPLC METHOD FOR THE DETERMINATION OF SPARFLOXACIN IN BULK AND PHARMACEUTICAL DOSAGE FORM

Authors

  • P. RAVISANKAR Vignan pharmacy college
  • CH. DEVADASU
  • G. DEVALA RAO
  • M. NAGESWARA RAO

Abstract

Objective: To develop a simple, accurate, precise and selective isocratic reversed phase-high performance liquid chromatography (RP-HPLC) method was developed for the quantitative estimation of Sparfloxacin in pharmaceutical formulations.

Experimental:  RP-HPLC method was developed by using Welchrom C18 Column with 250 mm x 4.6 mm i.d and 5µm particle size, SHIMADZU LC-20AT Prominence liquid chromatograph. The mobile phase used is phosphate buffer (pH-3.0): acetonitrile (70:30percent v/v) with a flow rate of 1mL/min.  The eluent was monitored by UV detection at 291nm using SHIMADZU SPD-20A prominence UV-Vis detector.

Results: The retention time of Sparfloxacin found to be 5.499 min. The method retains linearity in the range of 2-10µg/mL with correlation coefficient 0.9999. The mean recovery of sparfloxacin was 99.702percent to 100.456percent. The LOD and LOQ of the drug were found to be 0.186 µg/mL and 0.558 µg/mL respectively.  

Conclusion: The proposed method can be successfully used for routine determination of sparfloxacin in pharmaceutical formulations.

 

KEY WORDS:  Sparfloxacin, Isocratic RP-HPLC, Correlation coefficient , Method Validation.

Downloads

Download data is not yet available.

Author Biography

P. RAVISANKAR, Vignan pharmacy college

1 Department of Pharmaceutical Analysis and Quality Assurance, Vignan Pharmacy College, Vadlamudi, Guntur - 522213, A.P., India.

Associate Professor

Published

01-02-2014

How to Cite

RAVISANKAR, P., C. DEVADASU, G. D. RAO, and M. N. RAO. “A VALIDATED RP-HPLC METHOD FOR THE DETERMINATION OF SPARFLOXACIN IN BULK AND PHARMACEUTICAL DOSAGE FORM”. Asian Journal of Pharmaceutical and Clinical Research, vol. 7, no. 6, Feb. 2014, pp. 135-40, https://mail.innovareacademics.in/journals/index.php/ajpcr/article/view/848.