METHOD DEVELOPMENT AND VALIDATION OF LERCANIDIPINE IN HUMAN PLASMA BY LIQUID CHROMATOGRAPHY TANDEM-MASS SPECTROMETRY

Authors

  • Yahdiana Harahap Bioavailability and Bioequivalence Laboratory, Faculty of Pharmacy, Universitas Indonesia
  • Norma Andriyani Bioavailability and Bioequivalence Laboratory, Faculty of Pharmacy, Universitas Indonesia
  • Harmita . Bioavailability and Bioequivalence Laboratory, Faculty of Pharmacy, Universitas Indonesia

DOI:

https://doi.org/10.22159/ijap.2018v10i4.26544

Keywords:

Lercanidipine, Amlodipine, LC-MSMS, Liquid-liquid extraction, Validation

Abstract

Objective: To obtain an optimum and validated method for analyzing lercanidipine in plasma using Ultra Performance Liquid Chromatography of Tandem Mass Spectrometry (UPLC-MS/MS).

Methods: The separation was carried out using 1.7μm (2.1 x 100 mm) Waters AcquityTM UPLC C18 column, a mobile phase of the 0.1% formic acid-methanol mixture (20:80 v/v) with isocratic elution, 30 °C column temperature, 0.2 ml/min flow rate and amlodipine as an internal standard. Mass detection was performed with a positive XBL TQD type Electrospray Ionization (ESI) in Multiple Reaction Monitoring modes. Lercanidipine was detected at m/z value of 612.11>280.27 and amlodipine was detected at m/z value 409.1>238.15. The optimum sample preparation method was a liquid-liquid extraction using 5 ml of n-hexane-ethyl acetate (50:50 v/v), vortex mixed for 3 min, centrifuged at 4000 rpm for 20 min, evaporated with nitrogen at 50 °C for 30 min, and the residue was reconstituted with 100 μl of mobile phase.

Results: The method was linear in the range of 0.025-10 ng/ml with r ≥ 0.9986. Accuracy and precision within-run and between-run met the requirements with %diff and %CV, not exceeding ± 15% and not more than ± 20% for Lower Limit of Quantification (LLOQ) concentration.

Conclusion: It was concluded that the developed method met the requirements of selectivity, carry over, stability, the integrity of dilution, and matrix effects under the Guideline on Bioanalytical Method Validation by the European Medicines Agency in 2011.

 

Downloads

Download data is not yet available.

References

McClellan KJ, Jarvis B. Lercanidipine: a review of its use in hypertension. Adis Drug Evaluation; 2000. p. 1123-40.

Food and Drug Supervisory Agency of the Republic of Indonesia. Regulation of the Head of the National Agency of Drug and Food Control of the Republic of Indonesia Number HK.03.1.23.12.11.10217 Y 2011 on Drugs that are compulsory to take the equivalent test; 2011.

European Medicines Agency: Committee for Medicinal Products for Human Use (CHMP). Guideline on Bioanalytical Method Validation. London: European Medicines Agency; 2011.

Chaudhary DV, Patel DP, Shah PA, Shah JV, Sanyal M, Shrivastav PS. Determination of lercanidipine in human plasma by an improved UPLC-MS/MS method for a bioequivalence study. J Pharm Anal 2015;1-8. Doi:10.1016/j.jpha.2015.09.001

Younes KM, El Kady EF. A new validated bio-analytical liquid chromatographic-tandem mass spectrometric method for the quantification of lercanidipine in human plasma. Intl J Anal Bio Chem 2013;3:169-77.

Jabor V, Coelho E, Ifa D, Bonato P, Santos N, Lanchote V. Enantioselective determination of lercanidipine in human plasma for pharmacokinetic studies by normal-phase liquid-chromatography-tandem mass spectrometry. J Chromatogr B 2003;796:429-37.

Kalovidouris M, Michaela S, Robola N, Koutsopoulou M, Panderi I. Ultra performance liquid chromatography/tandem mass spectrometry method for determination of lercanidipine in human plasma. Rapid Commun Mass Spectrom 2006;20:2939-46.

Tiwari NR, Shah N, Bhalani V, Mahajan A. LC, MS, and LCMS/MS studies for characterization of degradation products of amlodipine. J Pharm Anal 2015;5:33-42.

Published

07-07-2018

How to Cite

Harahap, Y., Andriyani, N., & ., H. (2018). METHOD DEVELOPMENT AND VALIDATION OF LERCANIDIPINE IN HUMAN PLASMA BY LIQUID CHROMATOGRAPHY TANDEM-MASS SPECTROMETRY. International Journal of Applied Pharmaceutics, 10(4), 87–91. https://doi.org/10.22159/ijap.2018v10i4.26544

Issue

Section

Original Article(s)

Most read articles by the same author(s)

<< < 1 2 3 4 5 > >> 

Similar Articles

You may also start an advanced similarity search for this article.